Product type

Colorimeters

Colour is a perception produced by a light source, a surface and an observer together, and a colorimeter replaces all three with fixed standards so that a colour can be written down as numbers. That is what makes colour a specification: a target is agreed as coordinates, a tolerance is agreed as a difference, and from then on a sample either falls inside it or does not, regardless of who is looking at it or under which lamp.

What a colorimeter is

Colour is a perception produced by a light source, a surface and an observer together, and a colorimeter replaces all three with fixed standards so that a colour can be written down as numbers. That is what makes colour a specification: a target is agreed as coordinates, a tolerance is agreed as a difference, and from then on a sample either falls inside it or does not, regardless of who is looking at it or under which lamp.

The instrument works by illuminating the sample in a defined way and measuring the light that returns from it through a set of filters or a dispersive element, from which coordinates in a standard colour space are calculated. Two families are distinguished by their optics. A filter instrument, usually built around the forty-five degree geometry, illuminates at an angle and views at another, and reads a surface the way an observer standing beside it would. A spectral instrument measures across the wavelength range and computes the same coordinates from the spectrum, which allows it to do more with the data.

The distinction that decides most specifications is not the family but the geometry, and specifically whether the gloss of the surface is included in the measurement. A sphere, or integrating, geometry collects both the light reflected in the specular direction and the light scattered by the surface, so a sample is measured as though the gloss were part of its colour. A forty-five degree geometry excludes the specular component by design. The same paint measured under the two arrangements gives different numbers, and neither is wrong.

How a colorimeter measures

The measurement begins with the illuminant and the observer. A colour is only defined relative to a light source, so the instrument contains or assumes one, and a standard observer function stands in for the human response. Both are conventions agreed by international committee rather than properties of the instrument, and a reading quoted without them is not reproducible.

From the measured reflectance the instrument computes coordinates in a colour space, which describe the sample as a lightness and two chromatic components. A difference between a sample and a standard is then a distance in that space, expressed as a single number. The formula used to compute that distance matters, because different formulae weight the components differently and a pair of samples can sit on either side of a tolerance depending on which one is applied.

Repeatability and agreement are two different figures and are often quoted as one. Repeatability is the spread of readings taken on the same instrument of the same sample, and it sets the smallest change the instrument can detect. Agreement between two instruments is worse, because it depends on their individual calibration, their geometry and their spectral response. Two instruments can each be repeatable to a small fraction of a unit and disagree with each other by more than the tolerance being applied.

Configuration choices

Most of the specification is fixed by these decisions rather than by the accuracy class. Each one constrains the next, which is why they are settled in order.

ConfigurationOptions to fixWhat it actually decides
Measurement geometry Forty-five degree illumination with a normal view, or a sphere that includes the specular componentWhether gloss is part of the colour. A textured, grained or gloss-varied surface behaves differently in each, and the choice has to follow the sample and the tolerance.
Aperture size A small or large measurement area, and whether more than one is availableWhether a small feature can be measured without picking up its surroundings, and whether a textured sample is averaged over enough area to be representative.
Colour space and difference formula The coordinates reported, and the formula used to compute a difference between sample and standardWhether a quoted tolerance is being evaluated on the basis it was written for. The same pair of samples can pass under one formula and fail under another.
Illuminant and observer The light source assumed in the calculation, and the standard observer function appliedWhether a reading taken at one site can be compared with a reading taken at another. Without a common convention the numbers are not comparable at all.
Repeatability The spread of repeated readings on the same sample, and the conditions under which it is specifiedThe smallest difference the instrument can be relied on to detect, which has to be comfortably smaller than the tolerance it is being used to police.
Captured data and library Standards stored on the instrument, a colour library of measured samples, or a connection to a phone or computerWhether a judgement can be made at the machine and whether a record of it exists afterwards. A tolerance that is not recorded cannot be defended.
Calibration arrangement The white reference used to zero the instrument, and how often it is appliedWhether the readings stay comparable over a shift and between instruments. This is the practice that keeps a colour programme honest, more than the resolution of the display.

The geometry is the first decision, because it decides what the number means. Everything else refines how precisely that number is repeated.

How to choose a colorimeter

DecisionWhat settles itWhat to ask
The surface being measured Flat and uniform, or textured, grained and glossyTexture and gloss decide the geometry and the aperture together. A sample with a directional surface needs the arrangement that averages it rather than one that catches it at the wrong angle.
The size of the area to be judged A large panel, or a small printed detailAperture size decides whether the instrument measures the feature or the average of the feature and everything beside it.
Who sets the tolerance A customer specification, or a figure the producer choosesA tolerance quoted in a difference unit is only meaningful with the formula and the geometry attached; a tolerance without them is not a specification.
The margin the process can hold A wide tolerance, or one close to the repeatability of the instrumentWhere the tolerance is narrow, the repeatability of the instrument and the calibration practice become the limiting factors rather than the accuracy.
Whether the judgement is made on the line A decision at the machine, or a sample sent to a laboratoryA portable instrument puts the decision where the work is and needs a display, a battery life and a tolerance check built into it.
Whether colours have to be stored A standard or two, or a library of approved coloursA library makes the standard reproducible when a physical sample is lost or has faded, which is the usual reason a colour programme breaks down.
Whether the reading has to be reported A pass or fail marked on a traveller, or data exported to a systemThe export path is specified with the instrument, because a measurement that has to be retyped will eventually be retyped wrongly.

If any of these are still open, send what is known. A quotation can be issued against a duty description and refined once the gaps are filled.

The nearest alternatives

These are the types an enquiry for a colorimeter usually turns out to be, and the point at which the choice is genuinely decided.

Nearest alternativeWhat separates it from a colorimeter
Spectrophotometers Measure how much light a sample absorbs or transmits across a range of wavelengths, in order to determine what the sample contains. A colorimeter measures the colour of a surface against a standard. The instruments share an optical bench and answer different questions, and where a concentration is wanted the spectrophotometer is the one that is calibrated against a chemical standard.
TOC analyzers Report the total organic carbon in a water sample by oxidising it and measuring the result, which is a destructive determination of content rather than an assessment of appearance.
Water quality analyzers Report the condition of a water by several methods at once. Where colour is one of them, it is usually measured as absorbance in a cell rather than as the appearance of a surface, and the two meanings of colour should not be carried across.
Particle size analyzers Report a size distribution rather than a colour. A powdered sample is measured by both instruments and the two results describe different properties of the same powder.
Coating thickness gauges Measure how thick a coating is by a magnetic or eddy-current method. Colour and film thickness are the two figures a painting or coating specification usually sets together, and each needs its own instrument.

The boundary worth settling first is what the number is for: the appearance of a surface, or the concentration of a substance in a liquid. Both are called colour measurement and they are not interchangeable.

What to include in the enquiry

The first reply can be a quotation rather than a set of questions if the enquiry carries these. Where a figure is not known, send the duty description and we will confirm what it implies.

  1. The surface to be measured, including whether it is textured, glossy or grained
  2. The size of the area that has to be judged, and therefore the aperture required
  3. The tolerance to be applied, and the colour space and difference formula it is written in
  4. The illuminant and observer conventions the specification assumes
  5. Whether a portable instrument is wanted, or a bench instrument in a fixed location
  6. Whether a standard or a library of standards has to be stored and recalled
  7. Whether the readings have to be exported, and to what system
  8. The repeatability the process requires relative to the tolerance being policed
  9. Quantity, destination port and required delivery date

Available products

1 model in Colorimeters

A hand-held colorimeter standing upright in a white housing, with a measurement aperture at the base and an alignment triangle below it, a round control on the top face, and a printed label carrying a model designation, a reference number and a square code

LS170 (7 models) · Colorimeters

Portable Colorimeter

Handheld colorimeter that reads the colour of a surface and reports its difference from a stored standard, with a choice of measurement geometry, a colour library held on the instrument and a tolerance that returns a pass or fail.

Measurement time
About 1 second per reading
Illumination
Full-spectrum LED source, evaluated at illuminant D65 with a 10 degree observer
Measurement geometry
45/0 on the LS170, LS172, LS175 and LS372; D/8 with the specular component included on the LS171, LS173 and LS176
Measurement aperture
8 mm on the LS170, LS171, LS172, LS173 and LS176; 20 mm on the LS175; 4 mm on the LS372

Also in Analytical Instruments

Other product types in this category

Every type we quote for under Analytical Instruments. A type with no link yet is still quoted for - it is described on the Analytical Instruments overview, and gets a page of its own as soon as a model is published against it.

Frequently asked

Colorimeters questions

What does the difference figure actually tell me?

It is the distance between the sample and the standard in a colour space, expressed as one number, and it says how far apart they are without saying in which direction or why. Two samples can sit at the same distance from a standard and look quite differently wrong, one being too dark and the other too yellow, so the individual components behind the single figure are what a production engineer actually adjusts against.

Which geometry do I need?

It follows from the surface and from what the tolerance means. A sphere collects the specular reflection as well as the scattered light, so the measurement includes the gloss, which is the right behaviour when a surface is meant to be judged as a whole. The forty-five degree arrangement excludes the specular component, so it follows the diffuse colour of the surface. A tolerance written for one arrangement cannot be applied under the other.

Why do two instruments disagree on the same sample?

Because repeatability and agreement are different things. Either instrument may repeat its own readings very closely while the pair differ, and the difference comes from their individual calibration, their geometry and how closely their response matches the standard observer. It is the reason agreed instruments are usually the same model, kept on the same calibration routine, when a tolerance is tight enough to be argued about.

What has to be confirmed before a quotation?

The surface and the area to be measured; the tolerance and the colour space and difference formula it is written in; the illuminant and observer conventions assumed by the specification; whether the instrument has to be portable or can sit on a bench; whether standards or a colour library have to be stored; how readings are exported; and the repeatability the process needs relative to the tolerance being applied.

Quoting for Colorimeters

Send the specification and we will confirm the matching model, availability and lead time. Minimum order quantity is 1 pc.

Request for quotation

Request a quotation - Colorimeters

Send the technical detail and we will confirm availability, lead time and terms. In-stock items ship within 48 hours; customized specifications typically take 2-10 days. Minimum order quantity is 1 pc.

The more of this you can give us, the fewer exchanges are needed before we can quote.
No file chosen
Datasheet, drawing, BOM or photograph. One file, up to 10 MB (PDF, Office, image or ZIP).

By submitting this form you agree that we may contact you about this enquiry. We do not sell or share enquiry data. See our privacy policy.

Items held in stock are dispatched within 48 hours of order confirmation. Customized specifications are produced to order and typically ship within 2 to 10 days. Transit time then depends on the shipping method and destination.

Request a Quote WhatsApp