Application

Water Quality Monitoring

Measuring pH, conductivity, turbidity, dissolved oxygen and chlorine in process water and effluent.

Why this measurement is difficult to get right

Water quality monitoring has two distinct modes that are frequently confused. Online analysers run continuously on a process stream and are judged on drift and maintenance interval, because the value has to be trustworthy between visits. Laboratory measurements are made on discrete samples under controlled conditions and are judged on accuracy and documentation. The same parameter, measured two ways, is two different instruments with two different service lives.

Where a discharge consent or an internal specification applies, the parameter and the reporting threshold are already fixed by somebody else. That turns the instrument into a compliance component rather than a free technical choice: the question stops being which measurement is best and becomes which instrument can hold the required performance unattended until the next maintenance window.

The third factor is what the sample does to the sensor. Fouling, scaling, biological growth and reagent consumption all set how often somebody has to attend to the instrument, and that interval determines the real running cost far more than the purchase price does. Access for cleaning is therefore a design decision, not an afterthought.

What has to be measured

  • pH and ORP for reaction, neutralisation and corrosion control.
  • Conductivity and total dissolved solids for demineralisation and rinse quality.
  • Turbidity and suspended solids for filtration performance and final effluent.
  • Dissolved oxygen and residual chlorine for biological treatment and disinfection.
  • Temperature, because pH and conductivity both depend on it and compensation has to be applied somewhere.

Which instrument measures what

MeasurementWhat the reading has to establishInstruments commonly specified
Continuous pH, ORP and conductivity That a process is being held inside its control band between maintenance visits, which is a drift requirement rather than a bench accuracy oneProcess analysis instruments
Discrete sample measurement in controlled conditions An accurate figure for a sample that can be prepared, documented and repeated - suitable for compliance records rather than controlLaboratory instruments
Turbidity and suspended solids online Whether filtration is performing, which depends on stability at low values rather than on rangeTurbidity analysers
Disinfection residual That the required residual is present at the point of measurement, where it decays quickly and cannot be sampled and carried awayChlorine analysers
Sample composition and elemental content What the other measurements can only indicate indirectly, and often the parameter a consent actually limitsAnalytical instruments

A consent or specification usually names the parameter and the threshold, which fixes the measurement before the instrument is chosen.

Narrowing the choice for a water quality programme

  1. Establish whether the value is used to control a process or to report compliance, because that decides between an online analyser and a laboratory meter.
  2. Find the consent, specification or internal standard that names the parameter and the threshold, if one exists.
  3. Confirm the measurement range needed around the threshold, since a wide range often costs resolution where it matters.
  4. Establish the sample condition - temperature, pressure, solids, bubbles - and whether conditioning is needed upstream.
  5. Decide how often somebody can reasonably attend the instrument, because that sets the drift and cleaning requirement.
  6. Check that the sensor can be reached and removed for cleaning without shutting the process down.
  7. Confirm how temperature compensation is applied, since pH and conductivity both depend on it.
  8. Decide how the reading is recorded and how long it has to be kept, which follows from whether it is evidence.

Where this application ends and the next begins

Adjacent applicationWhat separates it from water quality monitoring
Water flow measurement That work reports how much water passed, while this one reports what was in it, and the two are recorded by entirely different instrument families.
Tank & silo level monitoring Level gives a quantity in a vessel; it says nothing about composition, and it is judged on headspace conditions rather than on sample condition.
Product quality inspection Inspection judges a manufactured article, whereas this work judges a stream whose composition changes with time and with upstream events.
Energy efficiency testing That work accumulates consumption; water quality is a concentration, and it is measured continuously for compliance rather than summed.
Industrial test bench & calibration A calibration bench verifies the analyser itself, which is a separate activity from using it to monitor a process.

What to send with the enquiry

  • The parameters to be measured and the threshold for each.
  • Whether the result controls a process or reports compliance.
  • The source - process water, effluent, potable or a treatment stage.
  • Expected range, and the condition of the sample at the measuring point.
  • Temperature and pressure at the point of measurement.
  • Whether sample conditioning is available or has to be included.
  • How often the instrument can be attended for cleaning and calibration.
  • Whether readings must be logged, and for how long they must be kept.
  • Quantity required and the date it has to be in use.
  • Whether the customer expects calibration support or consumables as part of the supply.

Published products

Instruments for water quality monitoring

A bench-top electric drying oven with its door closed, showing a double glazed viewing window, a rotary latch, and a control panel at the left with a two-line digital display, indicator lamps and a green mains switch

101-0A / 101-1A / 101-2A / 101-3A / 101-4A / 101-5A (desktop) and 101-0AS / 101-1AS / 101-2AS / 101-3AS / 101-4AS (vertical); 202-0A and 202-0B; A = galvanised chamber, B = stainless-steel chamber · Laboratory Ovens

Electric Drying Oven

Electric drying oven with a chamber from 16 to 1000 litres, a working range from 5 degrees above ambient to 250 degrees Celsius, a galvanised or stainless-steel interior, forced air circulation on most models and a timed digital controller.

Temperature range
5 degrees above ambient to 250 degrees Celsius, with a stated ultimate limit of 300 degrees Celsius
Chamber volumes
16, 43, 70, 136, 225, 640 and 1000 litres
Interior
Galvanised steel on the A versions and stainless steel on the B versions; the two are the same size in the same volume
Convection
Bottom heating with an adjustable blower on the desktop range apart from the 202-0 pair; rear heating with a rear blower on the vertical range, twin blowers at 640 litres
A panel-mount controller in a black enclosure with a colour graphic display, showing a live measurement with a temperature and a current output value, six keys along the lower edge and two indicator lamps

FK-EC480 / SIN-TDS210 / SIN-EC-E1 / SIN-EC-E2 controllers with TDS-6012, TDS-7001, TDS-6013, TDS-8002, TDS-7002 or ADE3500 electrodes; a general-purpose DC2100 controller is also described · Conductivity Analyzers

Online Conductivity Analyzer

Panel-mount online conductivity analyzer covering ultrapure water to brine, with automatic or manual temperature compensation, isolated current outputs, relays and a serial port, and a choice of contact or inductive electrodes.

Range
Set by the electrode cell constant: about 0 to 20 microsiemens per centimetre at a constant of 0.01, up to 0 to 600 millisiemens per centimetre on the high specification controller
Accuracy
Stated as 2.0 class on the flagship and high specification controllers and as 1.5 percent of full scale on the compact controller
Temperature
minus 10.0 to 130.0 degrees Celsius, with automatic or manual compensation
Temperature elements
NTC10K, Pt1000, NTC30K or Pt100 depending on the controller, matched to the element fitted in the electrode
White benchtop spectrophotometer with a rounded body, a red sample compartment lid set into the top, a sloped keypad panel carrying a small blue display at the right, and lettering printed along the lower front face above a black knob

LC-V1200E / LC-V1200 / LC-V1200N / LC-V1200NC / LC-UV1200 / LC-UV1200N / LC-UV1200NC / LC-UV1200Pro / LC-UV1800 · Spectrophotometers

UV-Vis Spectrophotometer

Benchtop single beam and double beam spectrophotometers in nine models, spanning the visible range on tungsten lamps and the ultraviolet and visible range on tungsten plus xenon, with bandwidths from 5 nm down to 1.8 nm.

Series
V series for the visible range on a tungsten lamp; UV series for the ultraviolet and visible range on tungsten plus xenon
Models
Nine model codes across the two series, from LC-V1200E to LC-UV1800
Wavelength range
350-1050 nm on LC-V1200E; 320-1100 nm on the other visible models; 190-1100 nm on every UV model
Spectral bandwidth
5 nm, 4 nm, 2 nm and 1.8 nm, by model
Panel-mount process controller with a red front panel over a pale blue liquid crystal display reading 3.76 pH, five black buttons along the lower edge, and a glass electrode with a pale blue bulb lying below the case

MIK-pH8.5 / MIK-pH6.5 / MIK-pH163S · pH Analyzers

Online pH/ORP Controller

Panel-mount pH and ORP controllers in three tiers, from a 4.3 inch display with two isolated current outputs and three alarm contacts down to a 96 x 96 mm case with one output, all reading pH to ±0.02 from a single glass or antimony electrode.

Tiers and displays
4.3 inch display on the flagship; 2.8 inch on the upgrade and economy tiers
Case size
151 x 144 x 118 mm flagship; 100 x 100 x 150 mm upgrade; 96 x 96 x 112 mm economy
pH range
-(2.00~16.00) pH on the flagship, with antimony electrode support; 0~14 pH on the upgrade and economy tiers
ORP range
±2000 mV on the flagship; -1000~+1000 mV on the economy tier; -1000~+1000 mV on the upgrade tier with ±2000 mV made to order
A panel-mount dissolved oxygen analyzer with a black bezel and a backlit liquid crystal display showing an output setup screen, with four keys beneath the screen, alongside a stainless-steel electrode with a threaded mounting body and a cable

OHR-DO10 (membrane) and OHR-DO20 (optical) controllers, with the DO170 membrane electrode or the DO260 optical electrode · Dissolved Oxygen Analyzers

Online Dissolved Oxygen Analyzer

Online dissolved oxygen analyzer for continuous water monitoring, with a membrane or an optical electrode, temperature, salinity and pressure compensation, two alarm relays, an analogue output and a serial port.

Measurement range, membrane controller
0 to 200 percent saturation and 0 to 20.00 mg/L or ppm, with a temperature channel from 0.0 to 65.0 degrees Celsius
Measurement range, optical controller
0 to 20.00 mg/L or ppm, with the saturation range carried as a reserved parameter; electrode temperature 0 to 60 degrees Celsius
Accuracy
plus or minus 1.5 percent of full scale on the membrane controller; plus or minus 0.5 mg/L on the optical controller; temperature to plus or minus 0.5 degrees Celsius
Compensation
Temperature automatically by a sensing element in the electrode or manually; salinity entered manually; pressure entered manually
Hand-held pen shaped pH meter with a light grey body, a dark grey cap at the top and a tapered probe at the foot, its display showing 7.49 pH above a reading of 30.1 degrees Celsius, with three buttons below the screen

PH848 / PH838 / PH818 / PH828 / PH828+ · pH Meters

Pen Type pH Meter

Pen type pH meters in five models across two accuracy grades, with the electrode and the display in one hand-held body, automatic temperature compensation and a selectable probe form.

Models
PH848, PH838, PH818, PH828 and PH828+ across two accuracy grades
pH range
0.0~14.0 on PH848; 0.00~14.00 on PH838, PH818, PH828 and PH828+
Resolution and accuracy
0.1 resolution at ±0.1 pH on PH848; 0.01 resolution at ±0.05 pH on the other four
Repeatability
0.1 on PH848; 0.03 on the other four

Frequently asked

Water Quality Monitoring questions

The questions that come up before an order. If yours is not here, send it with the duty and we will answer it alongside the quotation.

Should I choose an online analyser or a laboratory meter?

If the value is used to control a process, an online analyser is required, because a laboratory result arrives too late to act on. If the value is a periodic compliance or quality check on a sample, a laboratory meter is more accurate and far cheaper to operate.

Why is drift more important than accuracy for an online instrument?

Because nobody is watching it. A laboratory meter is calibrated and read in the same session, so a small accuracy advantage is realised. An online analyser has to remain inside its band unattended, so how far it moves between visits matters more than where it started.

Does a consent limit the instrument choice?

It fixes the parameter and the threshold, which in turn constrains the range and the accuracy class. What remains to be chosen is how the instrument holds that performance unattended, which is the part a consent does not specify.

Why does temperature compensation come up so often?

Because both pH and conductivity are temperature-dependent quantities. Compensation has to be applied somewhere, either in the sensor, in the transmitter or in the report, and where it is applied changes what the displayed figure actually means.

What drives the running cost?

Sensor life and cleaning interval, far more than purchase price. Fouling, scaling and biological growth decide how often somebody attends, and an instrument that is hard to reach will be attended less often than it needs and will then be distrusted.

Is sample conditioning part of the instrument supply?

It is a separate item, and it is worth deciding early whether it is in scope. Many installations discover after commissioning that the sample had to be cooled, degassed or filtered before the analyser could read it, and that work is easier to plan than to retrofit.

Quoting for water quality monitoring

Send the measurand, the range and the conditions, and we will confirm the instrument family, the configuration and the lead time.

Request for quotation

Request a quotation - Water Quality Monitoring

Send the technical detail and we will confirm availability, lead time and terms. In-stock items ship within 48 hours; customized specifications typically take 2-10 days. Minimum order quantity is 1 pc.

The more of this you can give us, the fewer exchanges are needed before we can quote.
No file chosen
Datasheet, drawing, BOM or photograph. One file, up to 10 MB (PDF, Office, image or ZIP).

By submitting this form you agree that we may contact you about this enquiry. We do not sell or share enquiry data. See our privacy policy.

Items held in stock are dispatched within 48 hours of order confirmation. Customized specifications are produced to order and typically ship within 2 to 10 days. Transit time then depends on the shipping method and destination.

Request a Quote WhatsApp