Product type

Portable Gas Detectors

A portable gas detector is worn or carried by a person to warn them of a hazardous atmosphere before it affects them. Its purpose is protection in the moment, which distinguishes it from a fixed detection system that protects a plant: a portable instrument has to survive being dropped, run on a battery, and sound an alarm that the wearer cannot miss.

What a portable gas detector is

A portable gas detector is worn or carried by a person to warn them of a hazardous atmosphere before it affects them. Its purpose is protection in the moment, which distinguishes it from a fixed detection system that protects a plant: a portable instrument has to survive being dropped, run on a battery, and sound an alarm that the wearer cannot miss.

The duties it covers fall into a few recognisable groups. Oxygen deficiency and enrichment, flammable gas or vapour expressed as a fraction of the lower explosive limit, and toxic gases such as carbon monoxide and hydrogen sulphide form the classical four-gas combination for confined space entry. Other duties - a specific solvent vapour, carbon dioxide in a fermentation vessel, a refrigerant leak - call for a different sensor set, and the choice follows from a risk assessment rather than from habit.

A pumped instrument with a sampling line allows the atmosphere to be tested before anyone enters, which is the sequence confined space work requires. A diffusion instrument responds only to the atmosphere at its sensor, so it protects the wearer while they are inside but cannot survey the space beforehand. Which of the two fits depends on the procedure, and that is often the most consequential decision in the specification.

How a portable gas detector measures

Each gas uses a sensor technology suited to it, and knowing which is present matters because the technologies fail in different ways. A catalytic sensor burns a small amount of the gas on a heated element and measures the heat released, which suits flammable gas and reads the fraction of the lower explosive limit directly - but it needs oxygen to work and is poisoned by silicones, leaded compounds and some solvents. An infrared sensor absorbs light at a wavelength the gas absorbs, needs no oxygen and resists poisoning, and it cannot respond to hydrogen or to some light hydrocarbons. An electrochemical cell reacts the gas at an electrode and produces a small current proportional to concentration, which suits toxic gases and oxygen; cells deplete with exposure and have a finite life. A photoionisation detector ionises volatile organic compounds with ultraviolet light and gives a total reading without identifying the compound.

The measurement of flammable gas is expressed as a fraction of the lower explosive limit, and that figure depends on the gas. The same instrument responds differently to propane than to the methane it may have been calibrated on, because the response varies between gases and the calibration holds only for the one it was set to. Where a specific vapour is the hazard, the instrument is calibrated to that vapour, or the reading is corrected for it.

Oxygen is measured for two opposite reasons. A deficiency in a confined space can render a person unconscious without warning, which is why oxygen is measured alongside the toxic and flammable channels rather than treated as an optional extra. An enrichment, which can follow a leaking oxygen line or rusting steel in an enclosed space, raises the flammability of everything present instead. Both conditions are reported on the same channel, and both matter.

Configuration choices

Most of the specification is fixed by these decisions rather than by the accuracy class. Each one constrains the next, which is why they are settled in order.

ConfigurationOptions to fixWhat it actually decides
Sensor set and channel count Single gas, or two, three or four channels in one instrumentWhich hazards are covered at once, and how much a person has to carry to be protected against all of them.
Sensor technology per channel Catalytic, infrared, electrochemical or photoionisation, chosen per gasWhether the sensor responds to the specific hazard, and whether it will survive the atmosphere it is exposed to.
Sampling method Diffusion, or a pump drawing the sample through a lineWhether the atmosphere can be tested before entry, which is what a confined space procedure normally requires.
Alarm arrangement Audible, visual and vibrating alarms, and whether the levels are adjustableWhether the warning is noticed in the conditions of use. A wearer in a noisy area or in heavy clothing may miss a sound and catch a vibration.
Battery and runtime The runtime per charge, and whether the battery is replaceableWhether the instrument covers a full shift, including the pre-entry sampling period.
Data logging Whether readings, alarms and calibration events are recordedWhether the record of an entry can be produced afterwards, which an investigation may require.
Durability and ingress protection The drop rating and the ingress protectionWhether the instrument survives the work, which for something dropped on a steel deck is the difference between a usable tool and a liability.
Calibration station Whether the instrument is calibrated in a docking station, and whether tests are recorded automaticallyHow much of the calibration discipline depends on a person remembering to do it, and how the evidence is stored.

How to choose a portable gas detector

FactorWhy it decides the answerHow to settle it before enquiring
Risk assessment of the space or the task The hazards present decide which gases have to be measured, not a standard packageList the actual potential hazards, including anything the work itself could release.
Whether entry is involved Confined space entry brings pre-entry testing and specific channel combinations with itState whether the work is a confined space entry, and whether pre-entry testing forms part of the procedure.
The specific gases A sensor is chosen per gas, and one technology will not cover a list of themName the gases, and say whether a flammable reading is wanted for a specific vapour or for any combustible.
Anything that would poison or disable a sensor Silicones, leaded compounds, hydrogen sulphide and solvents damage particular sensors permanentlyState what else may be present, beyond the gases that are to be measured.
Working conditions of the wearer Noise, protective clothing, glove use, heat and the length of the shift all affect whether an alarm worksDescribe the conditions, so the alarm and the runtime can be judged against them.
Record and procedure requirement Entry permits and investigations may require evidence of calibration and of readings takenState what has to be recorded, and for how long.
Local approval regime An approval for use in a hazardous area may be imposed on the instrument itselfState the classification, and whether an approval certificate is required.

If any of these are still open, send what is known. A quotation can be issued against a duty description and refined once the gaps are filled.

The nearest alternatives

These are the types an enquiry for a portable gas detector usually turns out to be, and the point at which the choice is genuinely decided.

Nearest alternativeWhat separates it from a portable gas detector
Fixed gas detectors Installed at a point where a gas would accumulate, wired into a control system and protecting a plant rather than an individual. They run continuously without attention and cover places a person would not carry an instrument into, but they cannot follow a wearer around a site.
Multi-gas detectors A portable instrument carrying several channels at once, which is the usual configuration for confined space entry. The distinction from a single-gas device is coverage: the single-gas unit protects against the one hazard it was bought for and gives no indication of the others.
Single-gas detectors Built around one hazard, usually as a small personal monitor worn throughout a shift. They are lighter and simpler, and they make no claim to cover a space in which more than one hazard is possible.
Gas leak detectors Used to locate the source of a leak rather than to protect a person. The instrument is moved along a pipe run or around a fitting and the reading is interpreted as a gradient towards the leak, which is a different pattern of use from a personal alarm.
Combustible gas detectors Concerned specifically with flammable gas and vapour, usually expressed as a fraction of the lower explosive limit. Where that is the only credible hazard, a dedicated instrument offers more options for that duty than a general four-gas monitor.
Oxygen gas detectors Single-purpose instruments for oxygen deficiency or enrichment, used where oxygen is the only credible hazard - a cryogenic area, or a space served by an inert gas line. They answer a question the four-gas combination also answers, but with a much narrower focus.

The decision that most often goes wrong is the sampling method. A diffusion instrument carried into a space cannot tell anyone what the atmosphere was before they entered it, and no amount of calibration changes that.

What to include in the enquiry

The first reply can be a quotation rather than a set of questions if the enquiry carries these. Where a figure is not known, send the duty description and we will confirm what it implies.

  1. The gases to be detected, from a risk assessment of the task or the space
  2. Whether the work involves confined space entry and pre-entry testing
  3. Whether a sample has to be drawn from a distance, or the instrument is worn
  4. What else may be present that could affect a sensor
  5. Shift length and the working conditions of the wearer
  6. Whether readings, alarms and calibrations must be recorded
  7. Area classification, and any approval required for the instrument
  8. Whether a calibration or docking station is wanted, and how many instruments it will serve
  9. Quantity, destination port and required delivery date

Available products

1 model in Portable Gas Detectors

Black hand-held gas detector with a blue display reading H2S 0 ppm, O2 20.3 %, CO 0 ppm and EX 0 %LEL, the words Gas Detector printed below the screen, three buttons beneath the panel and four grilles for the sensing elements in the lower body

BH-4 (2 models) · Portable Gas Detectors

Portable Gas Detector

Hand-held multi-gas and single-gas detectors with audible, visual and vibrating alarms, covering the four-gas confined space combination as standard and a long list of single gases to order.

Range
The BH-4 multi-gas instrument and the BH-90 single-gas instrument
Channels
Four on the BH-4, covering combustible gas, oxygen, hydrogen sulphide and carbon monoxide; one on BH-90
Display error
≤±5 %FS
Response time
T90 < 60 s

Also in Gas Detection & Safety Instruments

Other product types in this category

Every type we quote for under Gas Detection & Safety Instruments. A type with no link yet is still quoted for - it is described on the Gas Detection & Safety Instruments overview, and gets a page of its own as soon as a model is published against it.

Frequently asked

Portable Gas Detectors questions

What is the difference between a bump test and a calibration?

A bump test applies a known gas to confirm that the sensor responds and the alarm sounds, which is a functional check performed at the start of a period of use. A calibration establishes or adjusts the reading against a known concentration. A bump test does not correct an out-of-range sensor, and a calibration does not prove the alarm works. A practical regime does both, at intervals defined by the procedure.

Can one instrument detect any flammable gas?

It can detect that something flammable is present, but the reading depends on the gas. A catalytic sensor is calibrated against one gas and its response to another differs, so a figure expressed as a fraction of the lower explosive limit is only accurate for the gas it was set to. Where a specific vapour is the hazard, the instrument is calibrated to that vapour or the correction factor for it is applied.

Why is oxygen measured as well as the toxic and flammable gases?

Because a deficiency can incapacitate a person before any other channel reads anything, and an enrichment makes everything present more readily flammable. Both conditions are invisible to the other sensors, and a confined space procedure normally requires oxygen to be measured before entry and continuously while anyone is inside.

What shortens a sensor’s life?

Exposure to substances the technology cannot tolerate. A catalytic sensor is poisoned by silicones, leaded compounds and some solvents, and that poisoning is permanent. An electrochemical cell is consumed by the gas it measures, so its life shortens with exposure and it has a finite shelf life even unused. An infrared sensor is far more resistant, which is one reason it is chosen where the atmosphere is unpredictable.

Quoting for Portable Gas Detectors

Send the specification and we will confirm the matching model, availability and lead time. Minimum order quantity is 1 pc.

Request for quotation

Request a quotation - Portable Gas Detectors

Send the technical detail and we will confirm availability, lead time and terms. In-stock items ship within 48 hours; customized specifications typically take 2-10 days. Minimum order quantity is 1 pc.

The more of this you can give us, the fewer exchanges are needed before we can quote.
No file chosen
Datasheet, drawing, BOM or photograph. One file, up to 10 MB (PDF, Office, image or ZIP).

By submitting this form you agree that we may contact you about this enquiry. We do not sell or share enquiry data. See our privacy policy.

Items held in stock are dispatched within 48 hours of order confirmation. Customized specifications are produced to order and typically ship within 2 to 10 days. Transit time then depends on the shipping method and destination.

Request a Quote WhatsApp