- Test frequencies
- 100 Hz, 120 Hz, 1 kHz and 10 kHz; 100 kHz added on TH2822C and TH2822E
- Basic accuracy
- 0.25% on TH2822A and TH2822C; 0.1% on TH2822D and TH2822E
- Measured parameters
- L, C, R and Z as primary, with D, Q, phase angle and ESR as secondary; DC resistance on TH2822D and TH2822E
- Display ranges
- L 0.00 µH to 1000.0 H; C 0.00 pF to 20.000 mF; Z and R 0.0000 Ω to 10.000 MΩ
- Resolution
- 40,000 counts on the primary parameter; 0.0001 on D and Q
- Test signal level
- 0.6 Vrms on TH2822A and TH2822C; 0.3 Vrms, 0.6 Vrms or 1 Vrms on TH2822D and TH2822E, from a 100 Ω source
- Test speed
- 4 readings per second, or 1.5 readings per second
- Interface
- Mini-USB virtual serial port with the SCPI command set
- Body
- 90 × 190 × 40 mm, 350 g
A multimeter tells you that a component conducts. An LCR meter tells you what it looks like at a stated frequency, which is the question that decides whether a capacitor is still a capacitor and an inductor is still an inductor. This range puts that measurement in a hand-held case.
Product overview
The four models share one body, one display and one test interface. What separates them is a short list: how far the test frequency goes, whether DC resistance is measured, and how tight the basic accuracy is. The first two decide what can be tested; the third decides whether the result is worth recording.
That ordering matters when a model is chosen. A buyer who starts from accuracy usually ends up paying for a frequency they never use, while a buyer who starts from the component list usually finds that the entry model answers every question except one.
How the handheld LCR meter works
The instrument applies an alternating test signal to the component and measures the voltage across it and the current through it. Their ratio gives impedance, and the phase difference between them separates the resistive part from the reactive part. From those two quantities the instrument derives inductance or capacitance, and the secondary parameters that describe how far the component is from ideal.
Because the measurement is made at a frequency, the result belongs to that frequency. A capacitor measured at one frequency and compared with a datasheet value taken at another is not a comparison at all, which is why the frequency set is the first column of any selection table and the reason two models in this range exist with an extended top frequency.
The five-terminal slot is the other part of the method. Two of the terminals carry the current and two sense the voltage across the component, with the fifth guarding the loop; the sense pair draws almost no current, so the resistance of the leads and of the contact is left outside the measurement. The three banana sockets do the same job through a lead, with the guard brought out separately.
How to select the right handheld LCR meter
Start with the components. A bench that checks electrolytic capacitors and chokes for a drive will spend its time at the lower frequencies and will not miss the extended range. A bench that checks switching inductors, ferrites or small film capacitors will need the higher frequency, because that is where those parts are specified.
Then decide whether DC resistance is part of the test. Winding resistance, track resistance and contact resistance are DC quantities, and a meter without that function cannot report them however good its impedance accuracy is. If the inspection sheet has a DCR line, the choice narrows immediately.
Then look at the accuracy step. The difference between the two accuracy grades is small in absolute terms and matters mainly where a reading is compared against a tight limit, or where the same component is measured repeatedly and the spread between readings has to be smaller than the tolerance being judged.
Finally, consider how the readings leave the instrument. The serial port and command set allow a bench PC to log results, which is what turns a hand-held check into a record. If that is not needed, the port costs nothing and changes nothing.
Send the component list, the frequency at which each part is specified and the tolerance bands in use, and a model and clip set can be quoted against them.
Measurement limitations and selection considerations
The instrument measures the component, not the circuit. A capacitor measured in place carries the impedance of everything connected around it, and a reading taken without removing or isolating the part is a reading of the board.
The test signal level is low, by design, so that the instrument does not disturb the component. That also means the measurement is made under small-signal conditions: a component whose value depends on bias, such as a ceramic capacitor with a strong voltage coefficient, will read differently in the instrument than in the circuit it is meant for.
The test frequency set is fixed rather than continuous. Where a datasheet specifies a part at a frequency the instrument does not provide, the nearest available frequency is a different measurement, and the difference should be established before a limit is set on it.
Contact matters more than it appears. At low impedance the residual of the leads is comparable with the value being measured, which is why open and short calibration exists and why the four-terminal clips are worth having for surface-mount parts. A reading taken through two hand-held probes is only as good as the contact pressure behind them.
Every value below is taken from source documentation. Where a parameter is not stated in the available material it is marked as information required rather than estimated.