- Torque range
- 100 to 1000 N·m, 200 to 2000 N·m and 300 to 3000 N·m, one per size
- Accuracy
- ±1.0%
- Measurement modes
- TRACK, PEAK and PPEAK
- Units
- N·m, kgf·cm and lbf·in
- Display
- Colour LCD with touch input, fourth generation
- Impact counting
- Number of impacts within a set measuring time
- Peak reset
- Adjustable from 0.5 to 3 s, third generation and above
- Auto power off
- After five minutes without operation
- Continuous use
- 15 hours on a full charge
- Supplied accessories
- LK26009A and LK26010
An impact wrench does not deliver torque; it delivers a series of blows. The number a bench has to record is the height of those blows, and often the spread between them, which is a different measurement from the steady pull a torque wrench makes.
Product overview
The instrument is in two parts. A display unit carries the controls and the reading; a separate transducer sits in the torque path at the tool output and is joined to the display by a cable. On a heavy tool that is clamped in a fixture, that split is what makes the reading visible while the tool is running.
Four generations share the measuring engine and differ in presentation: a basic unit, a character LCD unit with a fold-out stand that props the housing at an angle on the bench, a high-brightness numeric display for bright or dim rooms, and a colour touch display that can also record the torque curve. The largest sizes of the fourth generation are the ones described here.
How the impact torque tester works
The transducer is bolted into the load path between the tool and a joint that is stiff enough not to absorb the blow. Each impact strains the transducer; the instrument samples the resulting signal fast enough to follow the rise and the decay of a single pulse, which is why the sampling rate, not the display, sets the ceiling on what the unit can resolve.
The three modes answer three different questions. TRACK follows the signal as it happens and is used for zeroing and for a slow, steady application. PEAK holds the highest value in the burst, which is the right reading for a hand tool or a clutch that slips at a set level. PPEAK holds the difference between the highest and the lowest point of the pulse train, which is what a pneumatic or battery impact tool produces when the joint keeps rebounding between blows and the instantaneous peak overstates what the fastener receives.
Peak reset is on a timer. After the set interval the held value clears, so a run of readings can be taken without touching the instrument between them.
How to select the right impact torque tester
Work from the joint outward. The torque the joint is specified to is the number that has to sit comfortably inside the scale; the tool’s capability only tells you that the scale must be large enough not to be over-ranged. A tool measured near the top of a range leaves no headroom for the scatter between blows.
Then decide what has to be recorded. A pass or fail bench needs the held value and the alarm; a bench investigating why a joint is drifting needs the curve, which is an added function on the colour display rather than a standard one.
Then match the mechanical interface. The transducer has to accept the output of the tool, and the accessories supplied with the series are listed by code only, so the adaptor for a particular drive is settled with the supplier rather than assumed from the tool size.
Finally, confirm the supply at the bench. The series is described both with a rechargeable pack and, for the fourth generation, with a mains supply, and which one arrives matters if the unit is to be moved between cells.
Measurement limitations and selection considerations
An impact reading is a dynamic measurement, and it depends on the joint as much as on the tool. Joint stiffness, thread condition, washer type and the number of blows all move the number, so a figure is only comparable with another figure taken on the same fixture. Recording the impact count alongside the torque is what makes two runs comparable.
Peak-to-peak is a spread, not a torque, and it should not be compared with a peak figure from another instrument. On a tool with poor repeatability the spread widens even when the mean output is unchanged, which is a useful signal but a different one.
Where the torque is applied slowly and held, as in the calibration of a torque wrench, this is the wrong instrument: a calibrator with a loading arm and a rigid reaction bracket measures that condition directly. Where the fastener is small and the tool is a screwdriver, a bench tester with a clamping chuck keeps the tool and the joint aligned, which matters more at low torque than the range does.
Every value below is taken from source documentation. Where a parameter is not stated in the available material it is marked as information required rather than estimated.