- Model
- JITAI-510 / JITAI-515 / JITAI-512 / JITAI-514
- Measuring range
- 0.8 to 600 mm on the JITAI-510 depending on probe; 1 to 300 mm for the through-coating model; 0.15 to 300 mm on the ultra-thin model; 0.75 to 160 mm on the electromagnetic model
- Resolution
- 0.01 mm on the JITAI-510, 515 and 514; 0.001 mm on the JITAI-512
- Accuracy
- ±0.05 mm between 0.8 and 10 mm and ±(0.01 + H/200) mm above 10 mm on the JITAI-510, where H is the measured thickness
- Sound velocity
- 1000 to 9999 m/s, set by material
- Display
- 128 × 64 LCD with a high-brightness EL backlight
- Measuring mode
- Pulse to first echo, described as R-B1, with single-point and two-point calibration against a 4 mm block
- Supply and runtime
- Two AA cells, stated as over 30 hours of operation
- Storage
- 40 thickness readings and 40 parameter sets on the JITAI-510
- Working temperature
- -20 to 40 °C on the JITAI-510
An ultrasonic thickness gauge answers a question that nothing else answers cheaply: how much metal is left. Pipe, tube, pressure vessels and plate are measured for wall loss from one side only, with the instrument on the outside and no need to open, cut or drain anything.
It is also one of the easiest measurements to get wrong in a way that looks right, because the number on the display depends on a velocity the operator set and on an echo that has to be the back wall rather than something in front of it.
What the instrument is
A probe and a body. The probe sends a pulse into the material, listens for the return from the far face, and the body times the interval and multiplies it by the sound velocity set for the material. Everything else - the display, the storage, the calibration - serves that arithmetic.
The range in this material is four instruments rather than four versions of one. The JITAI-510 is the general instrument. The JITAI-515 measures through a coating. The JITAI-512 is built for thin sections. The JITAI-514 is electromagnetic and needs no couplant at all. They share a probe family and a measurement principle and differ in what they can be pointed at.
Setting the velocity
The instrument measures a time, not a thickness, so the velocity is the conversion factor and getting it wrong scales the whole result. Steel is around 5920 m/s and aluminium around 6340 m/s, so a velocity left set for steel while measuring aluminium reads several percent low, and the error grows with the thickness.
Two things follow. First, the instrument’s velocity function - deriving the velocity from a sample of known thickness - is the reliable way to set it on an unfamiliar material, and it is worth having. Second, the table of velocities for 24 materials that comes with the instrument is described in the material itself as approximate, which is how it should be treated: as a starting point to be confirmed on a sample rather than a setting to be dialled in and trusted.
Probes, and the two reasons they change
The probe has to get sound into the material and the echo back out, and two properties of the work decide which probe does that.
The first is the material’s structure. Cast iron and other coarse-grain material scatter the beam, so the probes quoted for it are lower frequency and larger diameter - which is why the two N02 probes carry a separate, much shorter range for grey cast iron alongside their steel range. A probe that works well on rolled steel will under-read or fail entirely on cast iron.
The second is geometry. Thin walls and small radii need a small-diameter high-frequency probe, which is what the microcrystal types are for; the material gives minimum diameters and minimum wall thicknesses for each. Surfaces below 300 °C need the high-temperature probe.
Measuring through a coating
A painted or coated surface presents the standard instrument with an extra echo, and the coating echo arrives before the back-wall echo. Taking it for the back wall reports the coating thickness rather than the metal, and the reading looks entirely plausible.
The two models that answer this do it differently. The JITAI-515 uses an echo-echo mode that suppresses the coating return and times the first back wall, which is why its stated echo-echo range is far shorter than its pulse-echo range. The JITAI-514 avoids the problem by not needing a couplant at all: it is electromagnetic, measures with lift-off, and reads through the coating. Choosing between them is a question about the coating, the surface and whether anything can be applied to it.
Measurement limitations and selection considerations
The material carries several inconsistencies and all are recorded rather than resolved. The JITAI-510 range is given as 1 to 600 mm on the configuration card and 0.8 to 600 mm in the specification table; the dimensions differ between the table and two of the cards; the runtime is written two ways; and every probe frequency is printed as mHz where MHz is plainly meant.
The last of those is worth naming because it is not a judgement call: a 2 mHz probe would be inaudible and unusable, so the unit in the material is a typographical error. It is recorded as printed and read as MHz, with the discrepancy listed so that no one has to make that decision twice.
The limits that belong to the method are the ones to design around. The instrument needs a far face for the echo to come back from, so it cannot measure a coating thickness, and it cannot see a lamination inside the wall in its normal mode. It needs a couplant on the piezo-electric models, which rules out surfaces that cannot be wetted or that are too hot for the couplant supplied. And it needs the material to be uniform enough for a single velocity to apply, which is where coarse-grain material, laminates and dissimilar joints raise questions that the instrument cannot answer on its own.
Ordering
The items that fix the answer are the material and whether it is coarse-grained; the thickness range including the thinnest section; whether the surface is painted, coated or lined, which decides between a standard instrument and one of the two through-coating variants; whether a couplant can be applied; the surface temperature; the geometry, including the smallest radius and the smallest diameter to be measured; whether readings must be stored and reported; and the quantity of points to be measured per inspection. Where the material’s own figures conflict, the enquiry is the place to settle them - the range, the dimensions and the probe frequencies are all quoted two ways in the material behind this page. Quantity, destination port and required delivery date complete the enquiry.
Every value below is taken from source documentation. Where a parameter is not stated in the available material it is marked as information required rather than estimated.