Conductivity Analyzers

A conductivity analyzer reports the ability of water to carry an electric current, continuously, and that figure stands in for the total dissolved ion content without ever naming an ion. It is the standard way of watching a water treatment train: the reading falls as a demineraliser does its work, rises when a bed exhausts, and separates a clean condensate from a leaking one long before anything else would notice.

What a conductivity analyzer is

A conductivity analyzer reports the ability of water to carry an electric current, continuously, and that figure stands in for the total dissolved ion content without ever naming an ion. It is the standard way of watching a water treatment train: the reading falls as a demineraliser does its work, rises when a bed exhausts, and separates a clean condensate from a leaking one long before anything else would notice.

The analyzer is a pair, not a box. A controller on a panel holds the electronics, the display, the outputs and the relays; an electrode in the water does the measuring. The electrode is defined by a cell constant, which is the geometry term that maps its conductance to a conductivity, and by the material it is made from, which decides whether it survives the water and how quickly it fouls. Between them those two properties fix the range the pair can measure.

The reading is meaningless without its temperature. Conductivity rises with temperature in a way that is close to proportional and varies with what is dissolved, so instruments correct every reading to a reference temperature before displaying it, and the correction is not optional on a process where the water temperature moves. What makes one analyzer different from another is usually the accuracy of that correction, the range of electrode constants it can be told about, and whether it can also derive the related quantities - dissolved solids, resistivity or salinity - from the same measurement.

How a conductivity analyzer measures

The contact cell is the classical arrangement: two electrodes of fixed area and separation in contact with the water, with an alternating voltage applied so that the measurement does not polarise the electrodes. The cell constant is that geometry, and it is what allows one instrument to cover a range from ultrapure water to a concentrated brine by changing the electrode rather than the electronics. Low constants suit low conductivity, where the current is tiny and the water is close to an insulator.

A contact cell has two weaknesses at the extremes. In very clean water the signal is small enough that cable capacitance and fouling matter, and in dirty or concentrated water the electrodes foul and polarise. The inductive, or toroidal, cell answers the second case: two coils are encapsulated so that the water itself forms the coupling between them, and nothing metal touches the sample. It cannot measure pure water well, and it is the right choice for slurry, pulp and concentrated solutions where a contact cell would foul in days.

The temperature element is part of the measurement rather than a convenience. Instruments apply a compensation that brings every reading to a common reference, normally twenty-five degrees, and the coefficient used depends on the ions present. Where the composition is unknown, a general coefficient is applied and the residual error is accepted as part of the uncertainty. A reading quoted without its reference temperature and its coefficient is not comparable with a limit or with another instrument.

Configuration choices

Most of the specification is fixed by these decisions rather than by the accuracy class. Each one constrains the next, which is why they are settled in order.

ConfigurationOptions to fixWhat it actually decides
Cell type A contact cell with wetted electrodes, or an inductive cell that puts no metal in the waterWhether the instrument can measure ultrapure water at all, and whether it will survive a fouling or corrosive stream. This is the first decision.
Cell constant The geometry of the electrode, which fixes the band of conductivity the pair can resolveThe working range. A constant chosen for concentrated water is insensitive in pure water, and the other way round.
Electrode material Stainless steel, graphite, or an engineering plastic, with the body matched to the mediumWhether the electrode survives the chemistry and the temperature, and how often it has to be cleaned. A material that corrodes changes the cell constant as it does so.
Temperature element The type of sensor inside the electrode, and whether it matches what the controller expectsWhether the compensation can be applied at all. A controller and an electrode whose elements disagree cannot produce a corrected reading.
Derived parameters Conductivity alone, or conductivity with dissolved solids, resistivity or salinityWhether one instrument can serve a second purpose, such as a resistivity readout on a pure-water loop. The derived figures follow from an assumed factor rather than from a separate measurement.
Outputs and relays Analogue outputs, alarm and control relays, and whether a cleaning cycle can be sequencedWhether the reading controls the plant or only reports on it. A relay that drives a valve is a different requirement from one that drives a lamp.
Mounting and connection Panel, wall or pipe mounting for the controller; immersion, flow-through or insertion for the electrode; cable length and terminationWhether the electrode sits in the stream or in a bypass where it can be isolated. The installation decides how much of the maintenance is possible without shutting the process down.

The electrode is chosen before the controller. A sophisticated controller attached to the wrong cell constant has a narrower working range than a simple one attached to the right cell.

How to choose a conductivity analyzer

DecisionWhat settles itWhat to ask
The band of conductivity From ultrapure water at a fraction of a microsiemens to a concentrated brineThis alone selects between a contact and an inductive cell and fixes the cell constant. Everything else follows from it.
How dirty or corrosive the water is Clean condensate, cooling water, sewage, pulp or an acid streamFouling rate decides between a contact cell that is cleaned often and an inductive cell that tolerates the medium.
Whether the temperature moves A controlled loop, or a stream whose temperature follows the weatherA moving temperature makes the compensation input the accuracy-limiting factor rather than the cell.
What the reading drives A display for an operator, a control loop, or a compliance recordControl and compliance push the requirements towards outputs, relays and a documented calibration, not towards resolution.
Whether a second quantity is needed Dissolved solids, resistivity or salinity from the same probeA derived figure is convenient and carries the assumed conversion factor with it, which has to be stated if the reading is to be defended.
Where the electrode can be installed Directly in the pipe, in a bypass, or dipped into a tankA bypass is what makes calibration possible without stopping the process, and it is decided at design time rather than afterwards.
The environment and the cable run Indoor panel, outdoor cabinet, and the distance to the electrodeA long cable run affects a low-constant cell more than the instrument, and the termination has to suit the electrode.

If any of these are still open, send what is known. A quotation can be issued against a duty description and refined once the gaps are filled.

The nearest alternatives

These are the types an enquiry for a conductivity analyzer usually turns out to be, and the point at which the choice is genuinely decided.

Nearest alternativeWhat separates it from a conductivity analyzer
Process conductivity meters The same physical measurement presented as a panel or field instrument rather than as an analyzer feeding a control system. Where the reading has to drive the plant and be logged, the analyzer form is used; where an operator simply has to be able to see it, a meter is sufficient.
TDS meters Report total dissolved solids, which is a conductivity reading converted through an assumed factor rather than a separate measurement. The conversion holds for one water and drifts for another, which is why the raw conductivity is normally the figure to specify.
Salinity meters Report salt content on a scale that is defined for seawater. They are the right instrument for a marine or estuarine duty and the wrong one for an industrial water, where the relationship between conductivity and salt content is not the one the scale assumes.
pH analyzers Measure acidity independently of the ion content, so a neutral solution can be either pure or strongly saline. The two are complementary rather than alternative, and the same water quality limit often constrains both.
Dissolved oxygen analyzers Measure a dissolved gas rather than the ionic content. Conductivity is nevertheless relevant to them, because the salinity it implies is one of the inputs that converts their reading into a concentration.
Water quality analyzers Combine conductivity with other parameters in one instrument or one cabinet. Where conductivity is the controlled variable, a dedicated analyzer is used, and the multi-parameter instrument covers the rest of the picture.

The boundary worth settling first is whether the figure is wanted as an ionic concentration, as dissolved solids or as salinity, because the three are derived from the same measurement through different assumptions.

What to include in the enquiry

The first reply can be a quotation rather than a set of questions if the enquiry carries these. Where a figure is not known, send the duty description and we will confirm what it implies.

  1. The band of conductivity to be covered, with the lowest and highest values expected in normal operation
  2. The composition and temperature of the water, including any fouling or corrosive tendency
  3. Whether a contact or an inductive cell is required, and the cell constant if it is already fixed
  4. The temperature element fitted in the electrode, and whether it matches the controller
  5. Whether dissolved solids, resistivity or salinity have to be shown as well
  6. The analogue outputs, relays and any cleaning sequence the installation requires
  7. Whether the reading has to be logged or networked, and on which protocol
  8. The mounting arrangement for the controller and electrode, cable length, and whether a bypass is provided
  9. Quantity, destination port and required delivery date

Available products

1 model in Conductivity Analyzers

A panel-mount controller in a black enclosure with a colour graphic display, showing a live measurement with a temperature and a current output value, six keys along the lower edge and two indicator lamps

FK-EC480 (11 models) · Conductivity Analyzers

Online Conductivity Analyzer

Panel-mount online conductivity analyzer covering ultrapure water to brine, with automatic or manual temperature compensation, isolated current outputs, relays and a serial port, and a choice of contact or inductive electrodes.

Range
Set by the electrode cell constant: about 0 to 20 microsiemens per centimetre at a constant of 0.01, up to 0 to 600 millisiemens per centimetre on the high specification controller
Accuracy
Stated as 2.0 class on the flagship and high specification controllers and as 1.5 percent of full scale on the compact controller
Temperature
minus 10.0 to 130.0 degrees Celsius, with automatic or manual compensation
Temperature elements
NTC10K, Pt1000, NTC30K or Pt100 depending on the controller, matched to the element fitted in the electrode

Also in Process Analysis

Other product types in this category

Every type we quote for under Process Analysis. A type with no link yet is still quoted for - it is described on the Process Analysis overview, and gets a page of its own as soon as a model is published against it.

Frequently asked

Conductivity Analyzers questions

Why does conductivity always have to be temperature corrected?

Because the ability of water to carry a current depends on how freely its ions move, and ions move more freely as the water warms. A reading taken at thirty degrees is higher than the same water would give at ten, so without correcting to a common reference two readings cannot be compared with each other or with a limit. The instrument applies the correction continuously, and the coefficient it uses is a property of the ions present rather than of the instrument.

Contact cell or inductive cell?

The inductive cell puts no metal in the water, so it tolerates slurry, pulp, oily streams and concentrated solutions where a contact cell would foul or polarise. In exchange it cannot resolve very low conductivity, which is exactly where the contact cell with a small constant excels. The choice is therefore made by the water, and it is unusual for either to be right across a whole plant.

How is the cell constant chosen?

It is chosen so that the expected readings fall in the middle of the band the electrode can resolve. A cell with a small constant produces a larger signal in clean water, where the current would otherwise be too small to measure, and a large constant keeps the signal in range in a concentrated solution. Selecting the electrode for the extreme ends of the range rather than the normal operating band is a common and avoidable error.

What has to be confirmed before a quotation?

The band of conductivity to be covered; the water composition and its temperature variation; whether fouling rules out a contact cell; the cell constant if it is already established; whether dissolved solids, resistivity or salinity are also required; what the outputs and relays have to drive; whether the reading is logged or networked; and where the electrode can be mounted, with the cable length to the controller.

Quoting for Conductivity Analyzers

Send the specification and we will confirm the matching model, availability and lead time. Minimum order quantity is 1 pc.

Request for quotation

Request a quotation - Conductivity Analyzers

Send the technical detail and we will confirm availability, lead time and terms. In-stock items ship within 48 hours; customized specifications typically take 2-10 days. Minimum order quantity is 1 pc.

The more of this you can give us, the fewer exchanges are needed before we can quote.
No file chosen
Datasheet, drawing, BOM or photograph. One file, up to 10 MB (PDF, Office, image or ZIP).

By submitting this form you agree that we may contact you about this enquiry. We do not sell or share enquiry data. See our privacy policy.

Items held in stock are dispatched within 48 hours of order confirmation. Customized specifications are produced to order and typically ship within 2 to 10 days. Transit time then depends on the shipping method and destination.

Request a Quote WhatsApp