pH Analyzers

A pH analyzer measures the acidity or alkalinity of a process stream continuously, in place, and turns that reading into a signal a control system can act on. Acidity is not measured directly: the sensor develops a small voltage across a glass membrane whose response tracks the hydrogen ion activity of the solution, and the analyzer converts that millivolt signal into a pH value against a temperature-compensated calibration.

What a pH analyzer is

A pH analyzer measures the acidity or alkalinity of a process stream continuously, in place, and turns that reading into a signal a control system can act on. Acidity is not measured directly: the sensor develops a small voltage across a glass membrane whose response tracks the hydrogen ion activity of the solution, and the analyzer converts that millivolt signal into a pH value against a temperature-compensated calibration.

The practical difference between an analyzer and a pH meter is what happens after the reading. A meter is carried to a sample and read; an analyzer is installed in the line or the tank and stays there, holding a signal for a dosing pump, a controller or a recorder. That is why the type appears wherever pH has to be held rather than merely checked - effluent neutralisation, chemical dosing, fermentation, boiler and cooling water treatment, and food or beverage processes where a batch has to be proven inside a band.

Most of the difficulty in these installations is not in the electronics. It is in the electrode: whether it survives the medium, whether the reference junction stays open in a dirty or protein-rich stream, and whether anyone can reach it to clean and calibrate it. This page covers those decisions, and the boundary between a pH analyzer and the neighbouring measurements that are often mistaken for it.

How a pH analyzer measures

A glass measuring electrode and a reference electrode form a cell whose voltage depends on the hydrogen ion activity of the solution between them. The relationship is logarithmic, so one pH unit corresponds to a fixed change in millivolts rather than a linear one; the analyzer turns that millivolt difference into a pH reading using a calibration established against known buffer solutions.

Two facts follow from the construction. First, the measurement is a comparison, so the reference half of the cell is as important as the measuring half. A blocked or depleted reference junction shows up as a drifting or sluggish reading that no recalibration cures. Second, the glass membrane is a consumable. It ages, it dehydrates if it is allowed to dry out, and it is attacked by hydrofluoric acid and by strong caustic at elevated temperature. A pH reading that will not hold calibration is usually reporting the condition of the electrode rather than a fault in the transmitter.

Temperature changes the cell output and it also changes the chemistry of the process itself. Automatic temperature compensation corrects the electrode response to the measured temperature, but it cannot correct a process whose pH genuinely shifts with temperature. Those are two different things, and only one of them is a compensation problem.

Configuration choices

Most of the specification is fixed by these decisions rather than by the accuracy class. Each one constrains the next, which is why they are settled in order.

ConfigurationOptions to fixWhat it actually decides
Measuring electrode type Glass bulb, flat-surface glass, and non-glass solid-state or antimony bodiesWhether the sensor survives the medium and how much maintenance it needs. A flat surface resists abrasion and coating where a bulb would foul or break.
Reference system Single or double junction, in gel-filled, refillable or solid-state formHow the reference behaves in a dirty, sulphide-bearing or protein-rich stream. A double junction separates the reference from the sample, which is what keeps it alive in aggressive media.
Temperature compensation Automatic, using an integral or separate sensor, or manualWhether the readout tracks the electrode response as the process temperature moves. Without it, a correct electrode reports a wrong number whenever the temperature differs from the calibration point.
Mounting style Immersion, insertion, flow-through, and retractable or ball-valve fittingsWhether the sensor can be reached and withdrawn without draining the line. On a pressurised line this decides whether calibration is a routine task or a shutdown.
Transmitter arrangement Integral head-mounted, or a separate analyzer mounted at a convenient pointWhether the value is read where the pipe is, or where a person can see and operate it.
Outputs and control 4-20 mA, alarm and dosing relays, and digital bus optionsWhether the analyzer only reports, or also drives a dosing pump or a valve. Relay control is what allows a standalone neutralisation loop.
Cleaning arrangement Manual cleaning, or an automatic jet, brush or ultrasonic cleanerHow often someone has to open the line. On a fouling duty the cleaning method largely sets the maintenance cost.
Process connection and body material Fitting size, insertion depth, and the wetted body materialWhether the assembly bolts into the existing boss and survives the medium at process temperature and pressure.

How to choose a pH analyzer

FactorWhy it decides the answerHow to settle it before enquiring
pH range and the control band A process held tightly near a neutral point and one that swings across the full scale make different demands on the electrode and on the calibration routineState the expected range and, more importantly, the band that has to be held.
Process temperature It drives both the electrode choice and how long the sensor lastsGive the normal and maximum temperature, including any cleaning or sterilisation cycle.
Chemical environment Hydrofluoric acid, strong caustic, sulphides and organics each attack a different part of the cellList every medium the sensor will see, including cleaning agents.
Solids, fibres and coating tendency Fouling blocks the junction long before it damages the glassDescribe what is in the stream and whether the line already needs cleaning.
Pressure, and whether it varies The reference junction responds to pressure, and a vacuum or a swinging pressure can pull the reference into the sampleState line pressure and whether it is steady.
Access for maintenance An analyzer that cannot be reached is an analyzer that driftsSay whether the sensor can be withdrawn from a live line and how much space is available around it.
Area classification and washdown These change the housing and the fittings, not only the labelState the classification and whether the area is hosed down.

If any of these are still open, send what is known. A quotation can be issued against a duty description and refined once the gaps are filled.

The nearest alternatives

These are the types an enquiry for a pH analyzer usually turns out to be, and the point at which the choice is genuinely decided.

Nearest alternativeWhat separates it from a pH analyzer
Process pH meters The same electrode chemistry in a simpler assembly, intended for local indication rather than for a control loop. Where the reading has to drive dosing or be recorded continuously, the analyzer is the instrument; where someone simply needs to see the value at the tank, the meter is enough.
ORP analyzers Measures the oxidation-reduction potential of the solution in millivolts, which is what disinfection and some oxidation dosing loops actually need. It is not a proxy for pH: a chlorinated stream can hold a steady pH while its ORP moves, and the reverse is equally possible.
Conductivity analyzers Counts dissolved ions rather than hydrogen ions. It is far more robust in dirty streams and needs no reference junction, which is why it is preferred for rinse-water monitoring and for detecting contamination. It cannot tell an acid from a base.
Chlorine analyzers Measures residual disinfectant directly. pH and chlorine are related in the chemistry but not interchangeable in the loop: where the obligation is a free-chlorine residual, measuring pH and inferring chlorine does not demonstrate compliance.
TDS meters Reports total dissolved solids, usually derived from a conductivity reading. It answers how much is dissolved, not whether the solution is acid or alkaline, and it is unaffected by the balance between the two.
Water quality analyzers A multi-parameter assembly that may include a pH channel alongside others. It suits a monitoring station where several readings are wanted from one point; a dedicated analyzer is the better choice where one measurement has to be controlled tightly.

Four of these are frequently specified where pH was in fact the requirement, and two of them cannot answer the question at all. Settling that before the instrument is ordered is cheaper than discovering it after installation.

What to include in the enquiry

The first reply can be a quotation rather than a set of questions if the enquiry carries these. Where a figure is not known, send the duty description and we will confirm what it implies.

  1. Medium name and composition, including cleaning and sterilisation fluids
  2. Expected pH range, and the band that has to be held
  3. Normal, maximum and cleaning temperatures
  4. Line pressure, whether it is steady, and whether vacuum occurs
  5. Solids, fibres, oil or coating tendency in the stream
  6. Whether the reading must drive dosing, or only be displayed and recorded
  7. Mounting style possible at the point of measurement, and access for removal
  8. Output required, and what will read it
  9. Area classification and washdown requirement
  10. Quantity, destination port and required delivery date

Available products

1 model in pH Analyzers

Panel-mount process controller with a red front panel over a pale blue liquid crystal display reading 3.76 pH, five black buttons along the lower edge, and a glass electrode with a pale blue bulb lying below the case

MIK-pH8.5 (3 models) · pH Analyzers

Online pH/ORP Controller

Panel-mount pH and ORP controllers in three tiers, from a 4.3 inch display with two isolated current outputs and three alarm contacts down to a 96 x 96 mm case with one output, all reading pH to ±0.02 from a single glass or antimony electrode.

Tiers and displays
4.3 inch display on the flagship; 2.8 inch on the upgrade and economy tiers
Case size
151 x 144 x 118 mm flagship; 100 x 100 x 150 mm upgrade; 96 x 96 x 112 mm economy
pH range
-(2.00~16.00) pH on the flagship, with antimony electrode support; 0~14 pH on the upgrade and economy tiers
ORP range
±2000 mV on the flagship; -1000~+1000 mV on the economy tier; -1000~+1000 mV on the upgrade tier with ±2000 mV made to order

Also in Process Analysis

Other product types in this category

Every type we quote for under Process Analysis. A type with no link yet is still quoted for - it is described on the Process Analysis overview, and gets a page of its own as soon as a model is published against it.

Frequently asked

pH Analyzers questions

Why does the reading drift away from calibration?

Almost always because of the electrode rather than the transmitter. A coated glass membrane, a blocked reference junction, a dried-out bulb or an aged sensor all produce a reading that will not hold. Recalibrating an electrode in that condition moves the error rather than removing it. The sequence that works is to clean the electrode, check the reference, calibrate in fresh buffers, and replace the electrode when it will no longer pass.

Can a pH analyzer be used where hydrofluoric acid is present?

Not with a glass electrode. Hydrofluoric acid attacks the glass membrane and dissolves it, so the measurement fails and the sensor is destroyed. Those duties need a different sensor chemistry, and the concentration involved decides whether one is available at all. The acid content has to be stated at the enquiry for this reason.

What is the difference between pH and ORP, and which is needed for chlorine dosing?

pH measures hydrogen ion activity; ORP measures the tendency of the solution to oxidise or reduce, in millivolts. Chlorination control is normally an ORP duty, because the sanitising power of a chlorinated stream depends on more than its pH. Where both are regulated, both are measured, and neither substitutes for the other.

Does temperature compensation mean the reading is correct at any temperature?

It corrects the electrode response to the measured temperature. It does not correct a process whose pH genuinely changes with temperature, and it does not remove the requirement to calibrate. Where a specification is written at a stated temperature, the reading and the specification have to be taken at the same temperature to be comparable.

Quoting for pH Analyzers

Send the specification and we will confirm the matching model, availability and lead time. Minimum order quantity is 1 pc.

Request for quotation

Request a quotation - pH Analyzers

Send the technical detail and we will confirm availability, lead time and terms. In-stock items ship within 48 hours; customized specifications typically take 2-10 days. Minimum order quantity is 1 pc.

The more of this you can give us, the fewer exchanges are needed before we can quote.
No file chosen
Datasheet, drawing, BOM or photograph. One file, up to 10 MB (PDF, Office, image or ZIP).

By submitting this form you agree that we may contact you about this enquiry. We do not sell or share enquiry data. See our privacy policy.

Items held in stock are dispatched within 48 hours of order confirmation. Customized specifications are produced to order and typically ship within 2 to 10 days. Transit time then depends on the shipping method and destination.

Request a Quote WhatsApp