- Output current
- 3 to 33 A on 7305, 1 to 40 A on 7314, 1 to 60 A on 7316
- Maximum output voltage
- 6 V on 7305, 8 V on 7314, 12 V on 7316
- Resistance range
- 0 to 510 mΩ on 7305; 0 to 600 mΩ on 7314 and 7316
- Frequency
- 50 or 60 Hz, selectable, ±100 ppm
- Program memory
- Five groups on 7305, ten groups on 7314 and 7316
- Display
- 16 × 2 character LCD on 7305, 128 × 64 LCD on 7314 and 7316
- Test time
- Up to 999.9 s, continuous at zero
- Remote control
- PLC start, stop and reset inputs; pass, fail and processing outputs
- Standard accessory
- 7300 test lead set, listed for all three models
A ground bond test is a pass or fail judgement about a protective connection, and it is only worth repeating if the current, the hold time and the limit are the same on every unit. This range is built around that repetition, with output current as the main difference between the three models.
Product overview
Three models share one control layout. The smallest runs a bond test at a lower current and holds fewer programs; the two larger ones extend the current and add a graphic display with a longer program memory. All three apply an alternating current through the bond, read the resistance and compare it with an upper limit for a pass and a lower limit for a valid connection.
The front panel carries the start and reset buttons at the left, the display and four setup keys in the centre, and the current and return terminals at the right. The rear carries the mains voltage selector, the fuse holder, the earth stud and the two nine-pin connectors for remote control in and out.
How the ground bond tester works
A regulated alternating current is forced through the protective conductor and the point it is bonded to, and the voltage that appears across that path is measured at the same moment. The resistance follows from the two readings, and the instrument compares it with the limits the operator has entered. Above the upper limit the bond is too resistive and the test fails; below the lower limit the circuit is usually open, which is why a floor is set as well as a ceiling.
The sensed voltage includes the drop across the test leads and the clamps, and on a bond of a few milliohms that drop can be the larger part of the reading. The instrument therefore provides a self-zero correction: the lead resistance is measured and subtracted before the test, either automatically or as a manually entered offset.
Current is regulated rather than simply switched on, which is why the output is quoted as an accuracy on the setting instead of a band around a nominal value. A regulated current also means the reading does not drift as the bond heats and its resistance rises during the hold.
How to select the right tester
Start with the test current written into the product standard, not with the largest number in the table. That single figure decides between the three models, and choosing above it buys nothing except weight on the bench. The two larger models also allow the current to be set in fine steps, which matters when the standard names a current that does not fall on a coarse setting.
The second question is how many products the bench sees. Program memory is the practical limit on changeover: the smallest model holds fewer groups, so a bench that alternates between several products through a shift will spend time re-entering parameters that the larger models would have stored.
The third is the interface. All three accept a remote start and report pass or fail, and several instruments can be run in step from one controller, which is what a multi-point fixture needs. If the operator reads the value rather than the verdict, the graphic display of the larger models shows more at once.
The last is the site supply. The three models are marked for different input ranges and the selector on the rear only covers the voltages the model supports, so the supply has to be confirmed before the order rather than at commissioning.
Measurement limitations and selection considerations
The instrument measures the path that exists at the moment of the test. A bond that passes on a cold sample can fail after thermal cycling, and a screw that was tightened to torque on the bench can loosen in a vibrating assembly, so the test is a check on the sample in front of it rather than a prediction of service life.
Contact quality dominates the low end of the range. Probe placement, surface finish and the pressure of the clamp can move the result by more than the instrument error, and the self-zero correction removes the lead but not the contact. The same jig and the same probe position have to be used if readings are to be compared over time.
Current heats both the bond and the leads. The material does not state a duty cycle, so the rate at which tests can be repeated at the top of the current range has to be confirmed before a line is planned around the instrument.
Finally, this is a bond test and not a dielectric test. It says nothing about insulation strength, leakage current or power consumption, and a product that needs those checks needs a separate safety tester or a combined instrument rather than a higher current setting on this one.
Every value below is taken from source documentation. Where a parameter is not stated in the available material it is marked as information required rather than estimated.