- Basic accuracy
- ±0.2 % (0.2 class)
- Measurement mode
- AC on the AC series; AC, DC and RMS on the AC/DC series
- Voltage input
- 3 to 600 V AC; 3 to 600 V AC + DC on the AC/DC series
- Current ranges
- 5 mA to 20 A up to 50 mA to 200 A, according to series and rating
- Measurement set
- Voltage, current, active power, power factor, frequency, apparent power, reactive power, phase angle, average power and accumulated energy
- Display
- High-brightness LED, five digits, four windows
- Interface
- RS232 with Modbus RTU as standard; RS485 or relay output to order
- Supply
- 220 V, or 85 to 265 V AC / 100 to 300 V DC
- Panel cut-out
- 210 × 95 mm
- Body
- 300.2 × 215 × 100.6 mm, about 2.3 kg
A power meter is bought to answer what a load is doing rather than what a supply is capable of. The instrument reads voltage and current at the same instant and computes everything else from that pair, which is the reason it cannot be replaced by two separate readings taken one after the other.
Product overview
The range covers single-phase measurement across three series. The alternating-current series is built for general electrical verification, a second version is aimed at efficiency work where the interesting figure sits far below rated load, and the third adds direct-current measurement for adapters, chargers and DC supplies.
What separates them in use is not resolution but the two ends of the current span. The top of the range decides the largest unit that can be tested, and the bottom decides whether a standby figure is a measurement or only a number on a display. Both ends are quoted per rating rather than per instrument, which is why a quotation for this family comes back as a series and a rating rather than as a model number alone.
How the meter works
Voltage across the load and current through it are sampled together and converted against a single timebase. Multiplying the two point by point and averaging the result over a whole number of cycles gives active power directly, without assuming that the waveform is sinusoidal or that the load is linear. Power factor then follows as the ratio of active power to the product of the root-mean-square voltage and current, and apparent and reactive power come from the same three quantities.
That arithmetic is the reason the instrument exists in this form. A rectified average multiplied by a rectified average gives the right answer only on a clean sine wave; a switched-mode supply, a dimmer or a drive produces a waveform for which the two differ, and the error runs in one direction. An instrument that samples and multiplies does not care about the shape of the waveform, which is why the same instrument can be trusted on both a heater and a power supply.
How to select the right power meter
Three answers settle the selection. The first is whether the duty includes direct current: if an adapter, a charger or a DC supply is to be characterised, the alternating-current series cannot substitute and the decision is made. The second is the current span at both ends, taken from the largest unit to be tested and the smallest consumption figure the test standard names. The third is the nature of the load, because a drive or a switching supply requires a conversion that agrees with the readings taken around it.
The remaining choices are made once and then fixed. A low-current series is the difference between measuring standby consumption and merely displaying it. An energy total with elapsed time turns a run into a consumption figure. An isolated bus or a relay output has to be ordered with the instrument, since neither can be added afterwards.
Measurement limitations and selection considerations
This instrument characterises a load at a bench or a test position. It does not diagnose the waveform that produced the reading: harmonic content, distortion, unbalance and transients belong to a power quality analyser, which is a different instrument with a different purpose. Where the question is why a figure is what it is rather than what the figure is, the power meter is the wrong end of the diagnostic chain.
Two operational limits follow from the same construction. High currents pass through an external shunt, so the accuracy of the highest ratings depends on a component that may not be part of the instrument; and the basic class figure applies under calibrated conditions, with the error growing as the power factor falls. Both are reasons to state the duty in the enquiry rather than to select from the parameter table alone.
Every value below is taken from source documentation. Where a parameter is not stated in the available material it is marked as information required rather than estimated.