- Jaw opening
- 25 mm
- AC current
- 1 mA to 600 A on the two AC-only models; 10 mA to 600 A on the AC/DC model
- DC current
- 10 mA to 600 A on the AC/DC model only
- AC and DC voltage
- AC 1 mV to 600 V; DC 0.1 mV to 600 V
- Resistance
- To 20 MΩ, 40 MΩ or 60 MΩ according to model
- Capacitance
- 1 pF to 4000 µF on the middle model; 0.01 nF to 100 mF on the AC/DC model
- Display count
- 1999, 3999 and 6000 according to model
- Conversion
- True root mean square on the AC/DC model; not stated for the other two
- Safety rating
- CAT 600 V as printed on the panel; one detail image states CAT II 600 V
- Supply
- Two AAA cells on the two AC-only models
A clamp meter exists because some conductors cannot be interrupted. Where the plant is running and the current is the thing being checked, nothing can be inserted into the circuit, and measuring from outside the conductor is the only method available rather than simply a convenient one.
Product overview
The range covers three instruments built on two different sensors. Two of them read alternating current through a transformer jaw and differ by the additional functions they carry; the third replaces that jaw with a Hall-effect sensor so that direct current can be measured as well, and adds true root mean square conversion and a higher display count.
That split is the first thing to settle. The two alternating-current models share the 25 mm jaw, the non-contact detection, the torch and the automatic ranging, and the difference between them is capacitance, temperature, frequency and relative measurement. The third is a different instrument underneath, and it is the only one of the three that can be taken to a battery bank or a drive output.
How the jaw measures current
Enclosing a conductor in a magnetic circuit couples the field it produces into the instrument. In a transformer jaw the coupling is inductive, so only a changing field produces a signal and a steady direct current produces nothing at all. In a Hall-effect jaw the core is gapped and an element senses the static field directly, which is what allows a direct current to be measured, at the cost of greater sensitivity to where the conductor sits between the jaws and of an offset that has to be zeroed out.
The rest of the chain follows the usual decision about alternating signals. An instrument that averages the rectified waveform is calibrated on a sine wave and under-reports as the waveform departs from one, whereas an instrument that computes the heating equivalent agrees with the other instruments on the same installation. That matters on drives and switching supplies and not at all on a resistive heater.
How to select the right clamp meter
The first question is whether direct current is anywhere in the duty, because that decides between two sensor types rather than between two specification levels. The second is the conductor: 25 mm of jaw opening sets a hard limit on what can be measured, and a jaw that does not close fully reads low rather than failing.
The remaining choices are about what else the instrument has to do. Capacitance and temperature are what turn a current check into a winding or capacitor check without a second instrument. A starting-current capture is what makes a motor start visible. The safety category at the point of measurement is a separate question from all of these, and it is a limit on what the instrument survives rather than on how well it measures.
Measurement limitations and selection considerations
The method has a floor. A small current produces a small field, and at the bottom of the range the reading becomes unreliable in a way that no amount of display resolution fixes; a low alternating current is better measured by inserting an instrument into the circuit, and a low direct current needs the jaw zeroed before it means anything. Reading a cable that carries both the outgoing and the returning conductor cancels the field and gives a reading near zero, which is why only a single conductor is enclosed.
The instrument reports current and nothing more. Power, power factor and energy need voltage as well, taken at the same time, and a clamp meter that reads only current cannot answer a consumption question however large its range. Where the load is a drive or a switching supply, the conversion method becomes part of the specification, and an instrument without true root mean square conversion will disagree with the rest of the site measurements.
Every value below is taken from source documentation. Where a parameter is not stated in the available material it is marked as information required rather than estimated.