- Resolution grades
- 4½ digit at 21,000 or 51,000 counts; 5½ digit at 119,999 or 120,000 counts; 6½ digit at 1,199,999 counts
- DC voltage
- 200 mV to 1000 V on TH1941; 500 mV to 1000 V on TH1942; 100 mV to 1000 V on the 5½ and 6½ digit models
- AC voltage
- 200 mV to 750 V on TH1941; 500 mV to 750 V on TH1942; 100 mV to 750 V on the 5½ and 6½ digit models
- Tightest one-year DCV uncertainty
- 0.02% of reading + 0.008% of range on TH1942
- Resistance
- 200 Ω to 20 MΩ on TH1941; 500 Ω to 50 MΩ on TH1942; 10 Ω to 100 MΩ on the 5½ and 6½ digit models
- Reading rate
- 25 readings per second on TH1941 and TH1942; 1000 per second stated for the 5½ and 6½ digit models
- Display
- Dual vacuum fluorescent display, or a 4.3 inch colour LCD with an English menu option on TH1953 and TH1963
- Interfaces
- USB standard on TH1941; USB, GPIB and RS232C on TH1951; SCPI command set
- Body
- 235 × 105 × 360 mm at 2.4 or 2.5 kg; 235 × 105 × 320 mm at 2.7 kg on the colour LCD models
A bench multimeter earns its place by being the instrument the other instruments are checked against, or by being the one that runs unattended for a week. Those are different requirements, and this range separates them by resolution rather than by features.
Product overview
Seven codes across three grades. The four-and-a-half digit models are the production instruments: fast, simple, with a dual display that shows two quantities at once. The five-and-a-half digit models add resolution, four-wire resistance and, on one model, a colour display. The six-and-a-half digit model is the reference instrument, with graphical views and the tighter resolution that a calibration bench needs.
All of them share a bench body with a front input panel, a mains inlet and a communication port at the rear. The VFD models carry a green dual display; the colour models carry a single LCD with a menu that can be set to English.
The functions are the ones a bench actually uses: voltage and current in both forms, resistance in two and four wire, frequency and period, continuity and diode. Maths functions - percent, dB, dBm and a relative offset - sit on the same keys.
How the instrument measures
The input is ranged, amplified and then digitised by an integrating converter. Integration is what rejects mains hum: the conversion window is chosen so that a whole number of mains cycles falls inside it, and the interference averages to zero rather than being sampled into the result.
Alternating quantities pass through a true-RMS converter, which squares the signal, averages it and takes the square root. That is why the reading follows the heating effect of a distorted waveform rather than the average of its half-cycles, and why the stated bandwidth matters: outside it the converter no longer follows the input.
Resistance is measured by forcing a known current and reading the resulting voltage. In two-wire mode the same leads carry both, so their resistance is included. In four-wire mode a separate sense pair taps the voltage directly at the component, and the lead resistance drops out of the result - which is the only way a low-value resistor can be measured meaningfully on a bench.
The comparison function turns the reading into a decision: upper and lower limits are entered, and the instrument indicates pass or fail with a buzzer. That is what lets it run as a sorting station without a computer.
How to select the right model
Start with the tolerance. The digit count should be one step better than the tolerance being judged, so that the instrument’s own uncertainty is a small part of the decision rather than a large part of it. Everything else follows from that.
Then the interface. A stand-alone bench needs nothing but the front panel. A bench that is part of an automated system needs the port its software already speaks, and GPIB is not fitted to every model here.
Then the resistance work. If the smallest resistance to be measured is comparable with the lead resistance, four-wire mode is required, and that decides between the lower and higher models in the range.
Then the display. The dual VFD is quicker to read when two quantities matter at once; the colour LCD is easier when the instrument is set up once and left, and it carries the graphical views on the six-and-a-half digit model.
Finally the environment. A production bench that is never switched off and a calibration laboratory that is temperature controlled put different demands on an instrument, and the calibration interval should be agreed at the point of order rather than assumed.
Measurement limitations and selection considerations
Resolution is not accuracy. The digit count tells how finely the instrument displays a value; the one-year uncertainty tells how far that value can be from the true one, and the two are quoted separately for a reason. A six-and-a-half digit display with a wide uncertainty is a precise-looking wrong number.
The uncertainty figures are stated as a percentage of reading plus a percentage of range, and the second term dominates at the bottom of a range. Measuring a small signal on a high range therefore loses most of the accuracy the instrument is capable of. Manual ranging, or choosing the range that puts the reading near full scale, is what recovers it.
Bandwidth is a limit that does not announce itself. An AC reading is true-RMS only inside the stated bandwidth; above it the reading falls away without an error indication, so a switching waveform measured on a 100 kHz instrument can read low and look stable.
Four-wire resistance removes lead resistance but not thermal EMF. Where dissimilar metals meet at the terminals, a small thermal voltage adds to the reading, and the usual remedy is to take a reading in both polarities and average, or to use the instrument’s offset compensation.
Finally, the input terminals have a voltage rating and the nameplate carries a CAT rating. Neither should be read as permission to connect the instrument to a distribution circuit; a bench multimeter is a laboratory instrument, and measurements on live installation wiring belong to an instrument designed for that category.
Every value below is taken from source documentation. Where a parameter is not stated in the available material it is marked as information required rather than estimated.