Electrical Test & Measurement · Electrical Test Equipment

Gas Discharge Tube Characteristic Tester

Tester for ceramic gas discharge tubes and surge arrester parts, with 5 to 2000 V output, leakage measurement to 2200 µA and a HANDLER interface for sorting.

HG2536 ·

Key specifications

Output voltage
5 to 2000 V, continuously settable, ±1%
Voltage rise rate
100 V per second, ±1%
Voltage measurement accuracy
0.5% ±0.1 V
Leakage range
0.00 to 2200.0 µA, error within ±1%
Constant current accuracy
1000 µA ±5 µA and 100.0 µA ±1.0 µA
Test speed
25, 32 or 44 ms per single-parameter reading
Display
240 × 128 dot-matrix LCD, voltage shown as XXXX.X V
Interface
HANDLER (PLC) output for a sorting machine
Body
360 × 325 × 105 mm, about 6.8 kg

A gas discharge tube is a gap that is meant to stay open until the moment it is needed. Testing it means finding the voltage at which it closes, and checking that it does not leak before then. Both are high-voltage measurements, and the fixture matters as much as the instrument.

Product overview

The HG2536 is a benchtop tester for ceramic gas discharge tubes and surge arrester parts. It produces a settable output from 5 to 2000 V and measures the voltage across the tube and the current through it, in either constant-voltage or constant-current mode. The sorting result is available on a HANDLER connector at the rear.

The voltage is displayed to one decimal place on the high range, and the leakage range runs to 2200.0 µA. A varistor voltage range and a U1mA reference are also provided, which is what allows the leakage to be measured at a percentage of the breakdown value found on the same part.

The body is 360 × 325 × 105 mm at about 6.8 kg, with the display at the left of the front panel, the keypad at the right and the two test sockets along the lower edge.

How the gas discharge tube tester works

In constant-voltage mode the instrument applies the set voltage and reads the current. Below breakdown that current is the leakage of an open gap, which is very small; the reading is taken at a defined voltage rather than at whatever the source happens to deliver, which is what makes it comparable between parts.

In constant-current mode the instrument forces a current and reads the voltage. As the current rises the voltage across the tube climbs until the gap breaks down and the voltage collapses, and the point at which that happens is the breakdown voltage. Forcing the current rather than the voltage keeps the energy delivered to the tube bounded, which matters because a gas tube that is driven hard can be damaged by the test itself.

The voltage is raised at a defined rate of 100 V per second. The rate matters: a tube that breaks down at a given voltage when the voltage climbs slowly can break down at a different voltage when it is applied as a step, because the discharge needs time to develop. A stated rise rate is what makes a breakdown figure reproducible.

The test terminals take a four-wire Kelvin lead, so the voltage is sensed at the tube rather than at the instrument output, which keeps the lead and contact resistance out of a measurement taken at up to 2000 V.

How to select the right gas discharge tube tester

The first question is the breakdown voltage of the tube. It has to fall inside the output range with enough headroom to reach it at the stated rise rate, and not so far below the top of the range that resolution suffers. A 2000 V output covers the common surge arrester types; a tube that breaks down above that is outside this instrument.

The second is the leakage specification. The voltage at which leakage is measured, and the current that counts as a failure, both have to be inside the instrument’s range and above its noise. A limit expressed in nanoamps is not checkable on a microamp range.

The third is the test method the specification assumes. Some tube data sheets give a breakdown voltage at a defined rise rate; others give a U1mA value with a leakage limit at a percentage of it. The instrument supports both, and the one to use is the one the drawing states, not the one that is easier to set up.

The fourth is the sorting arrangement. The HANDLER output carries the result, so a line that sorts automatically needs the connector and a fixture that presents the tube repeatably. A bench that tests by hand needs neither.

Send the tube type, the breakdown voltage, the leakage limit and the test method the drawing assumes, and a configuration can be proposed against them.

Measurement limitations and selection considerations

The output is a DC high voltage. The instrument does not produce a surge, so it characterises the DC breakdown and the DC leakage; the impulse response of a gas tube, which is how it behaves in the event it is installed to protect against, is a different measurement and needs a different generator.

Contact quality decides the result at these voltages. A poor contact produces an arc at the terminal before the tube breaks down, and the instrument then reports the arc rather than the tube. This is why the output is settable rather than fixed: running just above the expected breakdown keeps the test energy low and the contact clean.

The tube’s history matters. A gas tube that has been struck repeatedly can break down at a lower voltage than a new one, and a tube measured immediately after a breakdown may need time to de-ionise before the next reading is meaningful. A fixture that tests the same tube twice in quick succession will not give the same two numbers.

The leakage of an open gap is small and the measurement is at high voltage, so cable movement and humidity both show up in the reading. A dry day and a humid one are not the same test.

Finally, the material gives a percentage-of-U1mA leakage condition without stating the default percentage or its range, and describes the sorting output without a bin count. Those are worth settling before the instrument is written into a test procedure.

Features

  • Output voltage is continuously settable from 5 to 2000 V, so the test can be run at the voltage that suits the tube rather than at a fixed step
  • Two working modes: constant voltage for reading leakage, and constant current for reading the voltage across the tube
  • Leakage can be measured either at the set voltage or at a percentage of the U1mA value found on the same part, which is how a varistor-style limit is applied
  • A leakage range to 2200.0 µA with an error within ±1%, so the reading is resolved where a gas tube's leakage actually sits
  • Single-parameter readings in 25, 32 or 44 ms, fast enough to sort tubes on a production line rather than sample them
  • A HANDLER (PLC) connector that carries the sorting result to automatic equipment without an operator
  • A 240 × 128 dot-matrix display that shows voltage and current together with mode, current step and speed

Technical specifications

Every value below is taken from source documentation. Where a parameter is not stated in the available material it is marked as information required rather than estimated.

High-voltage source
Output voltage5 to 2000 V, continuously settableStandard
Output voltage accuracy±1%Standard
Voltage rise rate(100 V ±1%) per secondStandard
Rise rate adjustmentThe material gives a single figure and does not state whether the rate is selectable or fixedInformation required
Constant voltage errorWithin ±0.5%Standard
Source impedanceAbout 2 kΩ, quoted on the varistor voltage measurement rowStandard
Measurement
ParametersVoltage and current, displayed as V and IStandard
Voltage displayXXXX.X V on the high-voltage rangeStandard
Voltage measurement accuracy0.5% ±0.1 VStandard
Varistor voltage range5.00 to 2000 V, used as the U1mA referenceStandard
Leakage measurement conditionAt the set voltage, or at the voltage obtained by multiplying the U1mA value by a set percentageStandard
Leakage percentage settingThe material does not give the default value or the adjustment rangeInformation required
Leakage range0.00 to 2200.0 µAStandard
Leakage errorWithin ±1%Standard
Constant current accuracy1000 µA ±5 µA and 100.0 µA ±1.0 µAStandard
Relation between the current steps and the leakage rangeThe material lists both but does not state how the range is switchedInformation required
Speed and sorting
Test speed, fast25 ms per single-parameter readingStandard
Test speed, medium32 ms per single-parameter readingStandard
Test speed, slow44 ms per single-parameter readingStandard
Sorting outputHANDLER (PLC) connector on the rear panel, carrying the sorting resultStandard
Sorting binsThe material describes the output but not the number of bins or how the limits are setInformation required
Display and controls
Display240 × 128 dot-matrix LCDStandard
Front-panel keysDisplay, setup, system, backspace and enter, trigger, digits 0 to 9, plus and minus, and four direction keysStandard
Test terminalsReference and test sockets on the front panel, for a four-wire Kelvin leadStandard
Electrical and environment
Supply220 V ±10%, 50 or 60 Hz ±5%Standard
Operating temperature0 °C to 40 °CStandard
Operating humidityBelow 90% RHStandard
Power consumptionNot stated in the materialInformation required
Ingress protectionNot stated in the materialInformation required
Mechanical
Dimensions (W × H × D)360 × 325 × 105 mm, measured by hand on the sampleStandard
WeightAbout 6.8 kgStandard
Supplied leadKelvin four-wire test leadStandard

Available models

Selection table
Model Configuration Key ranges Availability
HG2536 Gas discharge tube characteristic tester, benchtop, with HANDLER output 5 to 2000 V output; leakage 0.00 to 2200.0 µA; single-parameter reading in 25, 32 or 44 ms The only model in the material

Customization

  • A Kelvin test lead terminated for a particular tube holder or sorting fixture

Certification and compliance

Certification documents not yet supplied

Certifications are only published once a certificate copy can be produced, so this section is deliberately empty rather than populated with assumed approvals. Tell us which approvals your application requires and we will confirm what is available for this instrument.

Parameters still to be confirmed

The following are required for a complete specification but are not stated in the documentation supplied to us. They are listed rather than estimated:

  • Whether the voltage rise rate of 100 V per second is fixed or selectable, and its adjustment range if selectable
  • The default percentage and the adjustment range used when leakage is measured at a percentage of the U1mA value
  • The number of sorting bins and how the upper and lower limits are set, since the material describes the HANDLER output but not the sorting logic
  • How the two constant current steps of 1000 µA and 100.0 µA relate to the 0.00 to 2200.0 µA leakage range, and how the range is switched
  • The correct width of the body, printed as 360 mm on the dimension drawing and read as 330 mm from the parameter table during source transcription
  • Power consumption
  • Ingress protection of the enclosure
  • The complete packing list, of which the Kelvin lead, the mains lead and the fuse are named
  • Calibration certificate availability and standard
  • Warranty period and its terms

Frequently asked questions

What has to be confirmed before a quotation?

The DC breakdown voltage of the tube, since it has to fall inside the 5 to 2000 V output; the leakage limit and the voltage at which it is specified; whether the test has to run at a fixed voltage or at a percentage of a U1mA value measured on the same part; and whether the sorting result goes to a handler. The tube holder and the lead termination matter too, because a gas tube test at these voltages is decided by the fixture as much as by the instrument.

Why does the output voltage have to be settable rather than fixed?

Because a gas discharge tube breaks down at a voltage that is close to its rated value, and applying far more than that produces an arc at the contact rather than a clean breakdown. Setting the output just above the expected breakdown keeps the energy in the test low and makes the result repeatable. A fixed high output would trip the tube hard on every part.

Does the tester measure varistors as well?

The material gives a varistor voltage range and a U1mA reference, and the leakage can be measured at a percentage of that value. Those functions are described for this instrument. It does not, however, give a varistor-specific specification or a separate varistor model, so whether a particular varistor type is covered is a question for the supplier rather than something the material settles.

Related products

Other models in the same range

Light grey benchtop meter seen from the front, with three range keys and a power switch at the left, a four-digit red LED display in the centre showing 0.000, two red current terminals below it, and an INPUT socket with a COM terminal at the right beside two warning symbols

LK149A (5 models) · Electrical Safety Testers

Digital High Voltage Meter

Benchtop high voltage meter for verifying hipot tester output, covering 10 kV and 20 kV AC/DC ranges with 0.001 kV resolution and 0.4% DC accuracy.

Voltage ranges
0 to 10.000 kV AC and DC on LK149A, LK7005 and LK7015; 0 to 19.999 kV on LK149B and LK7020
DC accuracy
±(1% + 2 counts) on LK149A and LK149B; ±(0.4% of reading + 0.02% of range) on LK7005, LK7020 and LK7015
AC accuracy
±(1% + 2 counts) on LK149A and LK149B; ±(0.8% of reading + 0.02% or 0.04% of range) on the other three
Resolution
0.001 kV on every model
Bench capacitor leakage tester in a light grey case, with a backlit dot matrix display at the left showing a leakage current reading and charge status, function keys and a numeric keypad at the right, and a red high-voltage terminal beside two measurement terminals at the lower right

TH2689 (4 models) · LCR Meters

Capacitor Leakage Current Tester

Bench capacitor leakage testers with adjustable test voltage, leakage current measurement down to the nanoamp region, insulation resistance to 99.99 GΩ, timed charging and comparator limits.

Test voltage
1 V to 800 V on TH2689; 1 V to 500 V on TH2689A; 0 to 200 V and 0 to 500 V on TH2686N and TH2686C
Voltage accuracy
±(0.5% of setting + 0.2 V) on TH2689 and TH2689A; ±1% of setting + 1 digit on TH2686N
Leakage current range
0.001 µA to 20.00 mA on TH2689 and TH2689A; 10 nA to 19.99 mA with 1 nA resolution on TH2686N; 0 to 30 mA in 10 ranges on TH2686C
Insulation resistance
0.01 MΩ to 99.99 GΩ on TH2689 and TH2689A

Request a quotation for the HG2536

Send the configuration you need and we will confirm availability, lead time and terms. Minimum order quantity is 1 pc.

Request for quotation

Request a quotation - HG2536

Send the technical detail and we will confirm availability, lead time and terms. In-stock items ship within 48 hours; customized specifications typically take 2-10 days. Minimum order quantity is 1 pc.

The more of this you can give us, the fewer exchanges are needed before we can quote.
No file chosen
Datasheet, drawing, BOM or photograph. One file, up to 10 MB (PDF, Office, image or ZIP).

By submitting this form you agree that we may contact you about this enquiry. We do not sell or share enquiry data. See our privacy policy.

Items held in stock are dispatched within 48 hours of order confirmation. Customized specifications are produced to order and typically ship within 2 to 10 days. Transit time then depends on the shipping method and destination.

Request a Quote WhatsApp