- Model
- LDG-SUP
- Media
- Conductive liquids at ≥50 µS/cm, including water, wastewater, coolant, paper stock and acid or alkali solutions; ≥2 µS/cm for the liquid-cooling form and ≥5 µS/cm for the slurry form
- Nominal size
- DN10 to DN2000 standard, DN6 to DN65 micro-bore, DN300 to DN2200 to order
- Accuracy
- ±0.5 % of reading
- Range ratio
- 10:1
- Nominal pressure
- 1.6 MPa up to DN250, 1.0 MPa to DN1000, 0.6 MPa to DN2000
- Liner
- Neoprene, polyurethane, PTFE, PFA or F46, chosen against the medium and its temperature
- Electrode
- 316L stainless steel as standard, with corrosion-resistant alloy, titanium, tantalum and platinum-iridium options
- Display
- Instantaneous and total flow together on an integral display
- Remote reading
- An optional wireless module for reading the meter on a cloud platform
An electromagnetic flow meter is the instrument that measures a dirty liquid properly. There is nothing in the bore but the liquid itself, so nothing obstructs the flow and nothing catches fibres, grit or solids; the meter reads a voltage generated across the moving medium rather than the position of a moving part. That is why wastewater, slurry, paper stock and chemical dosing lines are its natural duties, and why it holds its accuracy on streams that would blind or jam other types.
The price of that construction is a precondition. The liquid has to conduct, and the pipe has to run full. Where either fails, no amount of instrument quality helps, and the answer is a different technology rather than a better magmeter.
What the meter does and what it needs
The measuring section is a lined length of pipe with coils around it and two electrodes through the liner into contact with the liquid. The coils generate a magnetic field across the bore; the moving conductive liquid generates a small voltage across that field; the electrodes pick the voltage up. Because the voltage is proportional to velocity for a given bore and field, the meter converts it into a volume per unit time.
Two consequences follow directly. A liquid below the conductivity the sensor needs cannot generate a usable signal, which is why deionised water, pure solvents and most hydrocarbon products are out of scope rather than merely difficult - and why this meter is offered with low-conductivity variants, at ≥2 µS/cm for cooling circuits and ≥5 µS/cm for slurry, when the standard threshold of ≥50 µS/cm is too high for the duty. And the electromagnetic circuit includes the pipe and the earth, so the grounding electrode the material describes is part of the measurement rather than an installation nicety: a meter without a solid earth reference holds a drifting zero, and no calibration will correct it.
Choosing the liner and the electrode
These two parts are where the medium compatibility of the whole instrument is decided, and they are best settled as a pair. The liner is what prevents the induced signal from short-circuiting through the metal body, and it is also the surface the medium touches for the life of the meter. The electrode is the one point where a metal is deliberately in contact with the liquid.
Two different failure modes have to be answered at once. Chemical attack decides the family: a fluoropolymer for aggressive acids and alkalis, a rubber for water and mildly aggressive service. Abrasion decides within the family: an abrasive slurry wears through a fluoropolymer at a rate no chemical argument anticipates, and the material offers rubber liners for exactly that reason. Temperature then narrows the choice again, because each liner carries its own medium limit, and the material gives these individually rather than as a single instrument figure - which means the liner, not the meter, sets the maximum temperature of a given order.
The material carries a separate selection table for each, with the corrosion behaviour of each electrode material described against the acid, alkali or salt it is expected to meet. Because the two tables are separate, the combination is worth confirming for the specific medium rather than assembling from two individually reasonable choices.
Reading the meter and getting the data out
The instrument shows instantaneous and total flow together on an integral display, which means a reading taken at the meter carries the shift total with it. For anything beyond local reading, the material describes an optional wireless module rather than a wired output: the meter reports to a cloud platform over a narrowband link, a mobile data terminal or an edge gateway, with multi-device monitoring and data export.
That is worth noting as a scope question rather than a feature, because a wired output is what a plant control system normally expects. Material received in November 2026 for the same range closes that gap: it states a 4-20 mA flow-current output, pulse or frequency output, and upper and lower alarm contacts, with RS485/Modbus on the communication form and a supply input of 220 V AC or 24 V DC. The two descriptions are kept side by side rather than merged, because they differ in detail - the pressure break points and the liner temperature limits are not identical between them - and the interface for a given order should be confirmed against the model actually quoted.
Installation
The material gives a straight run of 10 pipe diameters upstream and 5 downstream, and the reasons are the same as for any velocity-based meter: the profile at the electrodes is part of the measurement. It also gives the mounting positions that keep the pipe full and free of air - on the rising section of a line or the lower part of a horizontal run, never at the highest point of the line, never on a downward vertical - and advises a vent valve downstream where the fall in the line exceeds 5 m. The meter is not to be installed on the suction side of a pump.
The practical reason these rules are stated so firmly is that a magmeter that is badly installed does not fail. It reads, and it reads plausibly, and the error appears as a slow disagreement with a total taken elsewhere rather than as a fault anyone can point at.
Measurement limitations and selection considerations
The first limitation is conductivity, and it should be checked before anything else, because it decides whether the type applies at all. The second is that the pipe must run full: a partly filled line gives a reading based on a wetted area that is not the bore, and the error is largest when the flow is lowest, which is exactly when a plant is most likely to be looking for a leak or a blockage.
The material also leaves several figures unsettled, and they are listed as information required rather than resolved here. The most significant is the output: no wired analogue, pulse or serial signal is stated, and no supply voltage is given, so the electrical interface of an order is not yet determined by the material. The working temperature of -10 to 55 °C is given without saying whether it is the ambient or the medium temperature, and the maximum medium temperature appears only per liner. The dimension table in the material is reproduced with a note that its column meanings were inferred from the drawing. And the accuracy is given as a percentage of reading, which is not the same quantity as an accuracy class and should not be converted into one.
Ordering
The items that fix the answer are the medium, its conductivity and whether that varies; the minimum, normal and maximum flow; the pipe outside diameter, wall thickness and the flange standard it must match; the liner and electrode, or the chemical environment if those are not yet decided; the maximum medium temperature and the line pressure; the pressure rating required; the electrical interface and what will read it; whether the process is cleaned in place, buried or otherwise subject to conditions that change the body; and whether the area is classified as hazardous. A medium list and a pipe detail are enough for a covering quotation, which can then be refined as the liner and electrode are settled. Quantity, destination port and required delivery date complete the enquiry.
Every value below is taken from source documentation. Where a parameter is not stated in the available material it is marked as information required rather than estimated.