- Models
- A process loop version, a temperature-only version, and a multifunction version covering both
- Thermocouple types
- S, B, E, K, R, J, T and N, each with its own range and accuracy
- Resistance elements
- Pt100 and Cu50
- Resistance ranges
- 0 to 400 ohm, 0 to 1,000 ohm and 0 to 10 kilo-ohm
- Accuracy
- Stated as 0.05 percent of full scale on the measurement side for most types, with output accuracy stated separately and less tightly
- Cold junction
- Manual or automatic, with an external compensation input on the lower edge of the body
- Output modes
- Automatic step and ramp, as well as a steady value
- Display
- A colour screen able to show the measurement trend and the output trend at the same time
- Charging
- Type-C, from a phone charger
- Body
- 48 mm wide at the upper section, 132.2 mm high and 27 mm thick, with a silicone keypad
A temperature instrument on a plant is rarely somewhere convenient. It is welded into a vessel, mounted in a duct or buried under lagging, and the only place the loop can be reached is at the terminal head. A calibrator earns its place by being small enough to take there and by speaking the language of the sensor once it arrives, so that the loop can be checked where it stands rather than dismantled and brought to a laboratory.
That is a different job from the one a bath does, and the two are frequently confused. A bath provides a real temperature; a signal calibrator provides the electrical equivalent of one, and the difference decides what each of them can and cannot prove.
Product overview
The range covers three handheld models. One is built for the process loop, with direct voltage and current measurement and output, a transmitter output and a loop supply. One is built for temperature alone, covering thermocouple types, resistance elements and resistance. The third carries both sets and adds frequency measurement and output.
They share a body, a colour screen, a rechargeable battery charged over Type-C and a set of five terminals distinguished by colour.
How the instrument produces and reads a temperature
For a thermocouple the instrument generates a small voltage corresponding, on the relevant reference table, to the chosen temperature. That correspondence is only complete once the temperature of the terminals is accounted for, because a thermocouple reports the difference between its measuring junction and its cold junction. The instrument therefore applies cold-junction compensation, either measuring the terminal temperature itself or accepting a reference from the external socket on the lower edge, and adds the equivalent voltage to what it sources.
For a resistance element the instrument simulates the element instead, producing the resistance that corresponds to the temperature. The same circuits work in reverse when a device is being read, so the instrument can both source a value and measure one, at accuracies that are stated separately and are not the same.
How to select the right model
Start with the instrument list rather than with the calibrator. If the work is temperature only, the temperature model covers it. If the work is a loop that has to be powered and read in current or voltage, the loop model is the one, and the temperature functions are not needed. Where both appear on the same list, the multifunction model saves carrying two instruments, at the cost of one instrument that has to be understood properly before it is used.
Then confirm the signal types actually installed. A calibrator covers a defined set of thermocouple letters and resistance elements, and a sensor outside that set cannot be checked with it at all, however accurate the instrument is on the types it does cover.
Measurement limitations and selection considerations
The most useful constraint in the material is procedural: measurement and sourcing are advised against at the same time, particularly when resistance is being produced, because the load shifts the value being sourced. A loop check is therefore a sequence rather than a simultaneous operation.
Two things the page does not claim. It makes no statement about ingress protection, which the material does not give for an instrument intended for plant use. And it does not treat the nominal accuracy figure as the accuracy on every range: the measurement and output tables differ, type by type, and the page states both rather than the more flattering one.
Every value below is taken from source documentation. Where a parameter is not stated in the available material it is marked as information required rather than estimated.